background image

8-48

IM AQ1210-01EN

Advanced Analysis (Waveform Analysis) Screen

  

Multi trace analysis

Up to four waveforms can be loaded and compared.

2 way trace analysis

The waveform measured from each end can be 

combined and compared.

Differential trace

You can display a waveform obtained by subtracting the values of one waveform 

from those of another waveform selected as the current trace.

Multi Trace Analysis

3.

  Tap the 

Multi Trace

 item to display the Multi Trace screen.

 

  

Displays the waveform information loaded into Trace1 to Trace4 

(wavelength, dynamic range, power value)

Overview display

See section 6.2.

Trace1 waveform display

Trace2 waveform display

8.5  Performing Advanced Analysis (Advanced Analysis)

Содержание AQ1210D

Страница 1: ...User s Manual AQ1210A AQ1215A AQ1210D AQ1210E AQ1215E AQ1215F AQ1216F OTDR Multi Field Tester IM AQ1210 01EN 3rd Edition...

Страница 2: ...Handling Precautions IM 739884 01EN This document explains the handling precautions of the battery pack AQ1210A AQ1215A AQ1210D AQ1210E AQ1215E AQ1215F AQ1216F OTDR Multi Field Tester User s Manual IM...

Страница 3: ...ii IM AQ1210 01EN Revisions 1st Edition April 2019 2nd Edition April 2019 3rd Edition October 2020...

Страница 4: ...e s curit pouvant pr venir de tels accidents ATTENTION Attire l attention sur des gestes ou des conditions susceptibles de provoquer des blessures l g res ou d endommager l instrument ou les donn es d...

Страница 5: ...ble Installation Completion Notification 3 7 3 4 Rerouting Work 3 10 Chapter 4 Performing Averaged Measurements 4 1 Measuring in TRACE Mode 4 1 4 2 Measuring in MAP Mode 4 5 4 3 Performing Pass Fail J...

Страница 6: ...Displaying the Data Management Buttons 9 3 9 4 Saving and Loading Data 9 5 9 5 Creating Report Files 9 10 9 6 Performing File Operations 9 14 Chapter 10 System Setup 10 1 Displaying the System Setup...

Страница 7: ...se to detect fault locations in optical fiber cables and monitor fault conditions transmission loss splice loss etc It is mainly used in the following optical fiber cable installation and maintenance...

Страница 8: ...n perform an averaged measurement and then automatically display the measured result events as icons on the screen In addition if pass fail judgment conditions are specified judgment results are displ...

Страница 9: ...l power and the like simultaneously with the OTDR measurement In a measurement that takes a certain time to complete such as in an OTDR averaged measurement measurement of other fibers can be executed...

Страница 10: ...roject and make measurements under the same conditions In addition the AQ7933 Emulation Software can be used to create projects and the project files can be loaded into this instrument List Cores are...

Страница 11: ...rument See the Getting Started Guide IM AQ1210 02EN Communication dongle section 10 3 WLAN LAN OTDR port 2 OTDR measurement wavelength 850 nm 1300 nm1625 nm 1650 nm Light source output OTDR port 1 OTD...

Страница 12: ...vents that are generated from connections or splice points but are buried in noise The instrument derives the measured data by averaging the specified number of optical pulse measurements or by averag...

Страница 13: ...ed losses or reflections are known as events Incident ray Backscatter Splice Optical fiber cable Connector Bend Open end Near end reflection Splice loss Approximation line Reflection caused by a conne...

Страница 14: ...pagation is known as backscatter Small material Incident ray Backscatter Optical fiber cable Splice Loss Because spliced sections of optical fiber cables have a great number of changes in the material...

Страница 15: ...the area represented by the pulse width between the two points on the waveform at the level that is 1 5 dB below the peak value Attenuation Dead Zone An area where because there is a large reflection...

Страница 16: ...the cursor Distance Splice Loss When you place four markers as shown in the following figure the splice loss from the measurement start position to the measurement end position is calculated and the v...

Страница 17: ...ing start position marker 2 Event s falling end position marker Splice Loss at Two Locations 4 Point Monitor Splice losses at two locations can be measured simultaneously using four markers A B Distan...

Страница 18: ...the screen Event Display In TRACE mode detected events are displayed on the measured waveform In MAP mode the types of events are displayed using icons along with the distances of each from the measur...

Страница 19: ...sted Event number Distance from the measurement start point S Splice loss Return loss Event type Positive splice loss Negative splice loss Reflection Bending loss macro bending Splitter insertion loss...

Страница 20: ...esult that is less than the specified value End of Optical Fiber The instrument recognizes that an end of fiber E event has occurred when it detects a measured result that is greater than the specifie...

Страница 21: ...umber of splits The number of splitter splits Setting a Reference Trace You can perform averaged measurements or real time measurements while retaining the previous waveform on the screen reference tr...

Страница 22: ...dgment threshold on each measurement item splice loss return loss dB km total loss When the measurement value does not exceed the threshold the corresponding event is indicated as pass When the measur...

Страница 23: ...re optical fiber cables Model Measurement Light Wavelength AQ1210A AQ1215A AQ1210D AQ1210E AQ1215E AQ1215F AQ1216F 650 nm Optical Power Meter The optical power meter feature can be applied to models w...

Страница 24: ...Face Inspection FST option You can use a fiber inspection probe recommended by YOKOGAWA to take a photograph that shows the state of a fiber end face You can display this photograph on the instrument...

Страница 25: ...feature Like the OTDR feature when measurements are completed you can select between MAP mode and TRACE mode Event Analysis Using MAP Mode When you select MAP mode various events on the optical fiber...

Страница 26: ...ations by performing optical pulse measurement using multiple different pulse widths for the same wavelength and combining the multiple waveforms on the screen The instrument automatically determines...

Страница 27: ...s 100 the folder for saving the project file and that for saving measurement result files will be divided The folder is automatically divided every 100 fibers The divided folder names will be the proj...

Страница 28: ...eatures from the main view screen of multi fiber measurement OTDR feature realtime OTDR feature average Optical power meter Fiber inspection probe Example of measuring core number 44 Saving Measuremen...

Страница 29: ...surement light from the opposing instrument or AQ1100 AQ1200A AQ1200B AQ1200C AQ1200E AQ1205A AQ1205E AQ1205F that it is connected to and measures the optical power AQ1210A AQ1215A AQ1210D AQ1210E AQ1...

Страница 30: ...information about the fiber under loss test you must specify a fiber cable from the multi core optical fiber cable to use for the signal transmission Connect one end of the optical fiber cable for si...

Страница 31: ...optical fiber cable for signal transmission AQ1210A AQ1215A AQ1210D AQ1210E AQ1215E AQ1215F AQ1216F AQ1210A AQ1215A AQ1210D AQ1210E AQ1215E AQ1215F AQ1216F Master this instrument Execute a multicore...

Страница 32: ...vent of the current trace the closest event will be considered part of the current trace event Waveforms that meet the following conditions can be combined Both have the same wavelength Both have the...

Страница 33: ...dex Index Section Analysis You can set two markers start point S and end point E to measure the return loss and total loss in the section that you have specified By setting reference point B you can c...

Страница 34: ...This file is useful when you want to save an SOR file at a site where optical fiber cables are being installed and load the file into another instrument at a different site Only SOR files waveform da...

Страница 35: ...single sheet in a PDF file You can choose which items to include in the report File Operations Copying and Deleting Files You can copy or delete all the files and folders in the internal memory or ex...

Страница 36: ...date and time All All the above setup conditions are reset to their factory defaults All the files in the USER folder of the internal memory are also deleted The files in the following folders are not...

Страница 37: ...a screen with the data display area set to waveform display appears Map When the instrument starts a screen with the data display area set to icon display appears Setup Info When the instrument start...

Страница 38: ...a new firmware in a USB memory device and connecting it to USB port type A Installing Options You can install additional option licenses sold separately for the AQ1210 For details on the available opt...

Страница 39: ...km 30km 40km 50km 100km Set the pulse width AQ1210A AQ1210E AQ1210D 1310 1550 nm AUTO 5ns 10ns 20ns 30ns 50ns 100ns 200ns 300ns 500ns 1 s 2 s 5 s 10 s 20 s AQ1215A AQ1215E AQ1215F AQ1216F AUTO 3ns 10n...

Страница 40: ...e data save format is fixed to waveform data SOR extension Select the destination drive internal memory USB memory This is displayed when the mode is set to Date in which case you can set it Save dest...

Страница 41: ...ally Cable Length Distance Range Cable Length Distance Range Unknown Auto 16 km to 24 km 30 km 0 m to 160 m 200 m 24 km to 40 km 50 km 160 m to 400 m 500 m 40 km to 80 km 100 km 400 m to 800 m 1 km 80...

Страница 42: ...25dB 25 00dB 10 s 20 s 0dB 2 50dB 5 00dB 7 50dB 10 00dB 12 50dB 15 00dB 17 50dB 20 00dB 22 25dB 25 00dB 27 50dB 30 00dB 1 The pulse width is 5 ns on the AQ1210A AQ1210E AQ1210D 1310 1550 nm and 3 ns o...

Страница 43: ...ion or the number of times is determined by the specified average unit If Auto is selected one of the options above is used Event Search Event search is a feature that automatically searches for losse...

Страница 44: ...l time measurement it is performed at a sample interval optimized within the instrument As such if you perform a provisional pulse measurement using real time measurement check the waveform and then p...

Страница 45: ...to 70dB Set the splice loss 0 01dB to 9 99dB Set the IOR backscatter level See page 2 10 Executes initialization Press to reset the analysis settings to their factory defaults ANALYSIS tab Set macro...

Страница 46: ...s Fail Judgments 5 Tap Pass Fail Judgment to display the Pass Fail Judgment screen Turns the pass fail judgment on or off You can select items when Pass Fail Judgment is set to ON Set the splice loss...

Страница 47: ...of splits Unknown 2 Branches 4 Branches 8 Branches 16 Branches 32 Branches 64 Branches Set the number of splits None 1 2 Includes the selected split in the judgment conditions Set the splitter loss t...

Страница 48: ...e width 40 00dB to 94 99dB Pulse width used as a base for the backscatter level Set the backscatter level base 1 s 1ns Setting the Splitter Loss 5 Tap Splitter Loss to display the Splitter Loss screen...

Страница 49: ...optical fiber cable Display This function displays the detected bending losses in the event list on the event screen when optical pulses are measured OFF The bending losses are not displayed in the e...

Страница 50: ...ible values can be obtained Disadvantage If a large reflection or splice loss is present in the section whose loss is being calculated those values are also included in the calculation so large errors...

Страница 51: ...B 1300 nm 44dB 74dB 1310 nm 50dB 80dB 1550 nm 52dB 82dB 1625 nm 53dB 83dB 1650 nm 53dB 83dB Index of Refraction IOR The instrument uses the index of refraction to calculate the distance If you do not...

Страница 52: ...completion notification Configure settings that indicate that the optical fiber cable being installed has reached the destination with an alarm or the like Set the zoom method Natural Legacy Select t...

Страница 53: ...imation line of the event marker Y2 to the measurement point For each type the loss measurement range varies depending on whether the distance reference is set top illustration in the screen or not se...

Страница 54: ...PIN the default value is 0000 Enter the new PIN Enter the new PIN again for confirmation Confirms the input Configure the work completion notification 5 Tap Work Completion Notice Setup to display th...

Страница 55: ...ns on how to use markers see section 6 1 Example of 4 Point Markers Marker 1 2 3 Line After you select the line marker that you want to operate from a set of multiple line markers move the selected li...

Страница 56: ...play the approximation lines that are used to calculate splice loss or return loss OFF Approximation lines are not displayed ON Approximation lines are displayed Example of 4 Point Markers Marker 1 2...

Страница 57: ...ns Are Generated The optical pulse that is generated from position I in the following figure propagates in the direction of II Light ray A that is reflected at connection II in the figure is reflected...

Страница 58: ...t display area of measurement data to either of the following settings Return loss The ratio of the incident optical power level and the reflected optical power level is displayed Reflection level The...

Страница 59: ...loss between two points Measurement reference point S Note If Loss between S and E is selected markers 1 and 2 are placed automatically at the measurement reference point S and the END point E when e...

Страница 60: ...method does not include the near end reflection in the cumulate loss value The distance reference is set to the start point of the near end reflection The loss from near end reflection end point Y2 to...

Страница 61: ...Mode1 Events detected at positions other than the master event positions are also displayed in the event list Events other than fixed events are also detected An earlier event is detected so the numb...

Страница 62: ...ion VLS is installed The light source or visible light source ON soft key The LS Power Adjust soft key in an auto loss test LS loopback test The LS Power Adjust soft key in an multi core loss test LS...

Страница 63: ...lect the cursor display format from the following CROSS The position on the waveform is displayed with vertical and horizontal cursors that intersect LINE The position on the waveform is displayed wit...

Страница 64: ...Press SETUP to display the setup screen 4 Tap the File tab to display the following screen Set the report format See page 2 25 Set the file name See page 2 24 Set the file report format See page 2 26...

Страница 65: ...a b 2 a c 3 a d 4 a e 5 a f 6 a g 7 a h 8 Set the tape number start character Set within the specified tape number type range the tape number start character that will be attached to the file name Set...

Страница 66: ...FF ON Waveform display OFF ON Marker information display OFF ON Measurement condition display OFF ON Overview display OFF ON Event search condition display OFF ON Pass fail condition display OFF ON Ev...

Страница 67: ...to display 1Trace 2Trace Select the number of waveforms to show in a single page You can select the following items only when 2Trace is selected Order File Name Date ID Select the file export order by...

Страница 68: ...n Entering Text in the Getting Started Guide IM AQ1210 02EN You can enter up to 36 characters Select the check box to use the file ID ID number as the fiber ID For details on ID numbers see Setting th...

Страница 69: ...specified for field 1 under Name Type You cannot set duplicate items for Name Type Types of Characters and Strings Allowed in File and Folder Names The following restrictions apply to the types of cha...

Страница 70: ...and time1 2 Firmware version Information such as label and company name can be specified on the Label Setup screen explained in Setting Labels Set these if necessary Measured Results Link Summary The...

Страница 71: ...int Distance scale value Icon display Distance from the start point Distance between events Connector return loss Event Search Conditions The following information is included Splice loss Return loss...

Страница 72: ...e case of alphabet characters the order is a to z Z A In the case of numbers the order is 9 to 0 In the case of alphabet characters the order is z to a EventList Setup Set the items to show in the eve...

Страница 73: ...Procedure Real Time Measurement Screen OTDR screen 1 Press MENU to display the MENU screen For details on the MENU screen see the Getting Started Guide IM AQ1210 02EN 2 Tap OTDR to display the OTDR s...

Страница 74: ...ion see the explanation in section 10 1 If distance range and pulse width are set to auto the instrument automatically selects the optimal values internally immediately before starting a real time mea...

Страница 75: ...splays the loss at the position were each marker is placed Clears markers When Real Time Measurement Is Not in Progress Marker operation See section 6 1 Event analysis control See section 5 1 Snapshot...

Страница 76: ...e the distance to the open end end point is selected When setting the pulse width manually set a pulse width appropriate for the distance range For details on the appropriate pulse width see section 2...

Страница 77: ...en automatically switches to TRACE mode Saving Waveforms during Real Time Measurement If you save waveforms during real time measurement waveform data at that point is saved in memory If real time mea...

Страница 78: ...ms the waveform See section 6 2 Distance from the start point to the cursor position on the waveform and the level at the cursor position Resets waveform zoom See section 6 2 Select the loss measureme...

Страница 79: ...inishing real time measurement Configure the work completion notification 2 Tap the Advanced Measurement soft key and then the Work Completion Notice soft key The Work Completion Notice screen appears...

Страница 80: ...position If work completion notification is turned on CheckPoint is displayed at the cursor Place the cursor to a position that is a little after the installation point Work point Set the completion c...

Страница 81: ...cursor position 50 km 100 km Event Completion notification cursor position during optical fiber cable extension CheckPoint CheckPoint Notification Turns on or off the work completion notice feature O...

Страница 82: ...est mise travers les ports de source lumineuse Ne pas d brancher les c bles de fibre optique connect s Des l sions oculaires peuvent tre caus es si le faisceau lumineux p n tre l il When installed fi...

Страница 83: ...ing REAL TIME again stops the measurement Soft Key Menu Work Trace Comparison 2 Tap the soft key menu display button when real time measurement is in progress During real time measurement Captures the...

Страница 84: ...Because markers are not captured separately record the connection loss value between A and B before the construction A B The captured waveform is displayed dimmed Measuring the Connection Loss betwee...

Страница 85: ...e construction is moved to the right side Waveform data after construction Waveform data before construction captured waveform Moved to the right By overlapping the waveform data before construction a...

Страница 86: ...e Averaged Measurement Screen OTDR screen TRACE mode 1 Press MENU to display the MENU screen For details on the MENU screen see the Getting Started Guide IM AQ1210 02EN 2 Tap OTDR to display the OTDR...

Страница 87: ...automatically event analysis is executed and the analysis results are displayed on the screen If you press AVG again during averaged measurement measurement is stopped Note You can select the initial...

Страница 88: ...n Progress Marker operation See section 6 1 Event analysis control See section 5 1 Snapshot Reference trace control See section 6 3 Section analysis control See section 6 4 Set work completion See sec...

Страница 89: ...to the open end end point is selected When setting the pulse width manually set a pulse width appropriate for the distance range For details on the appropriate pulse width see section 2 1 Average Coun...

Страница 90: ...s es si le faisceau lumineux p n tre l il Procedure Averaged Measurement Screen OTDR screen MAP mode 1 Press MENU to display the MENU screen For details on the MENU screen see the Getting Started Guid...

Страница 91: ...t to auto the instrument automatically selects the optimal values internally immediately before starting a measurement Change these values if necessary For details see section 2 1 Explanation Averaged...

Страница 92: ...idth For details on the attenuation value appropriate for the pulse width see section 2 1 Switching the Display between TRACE Mode and MAP Mode You can switch the data display screen TRACE mode The da...

Страница 93: ...you tap the value switches between ON and OFF The result of the pass fail judgment on the waveform data shown on the data display screen Pass fail judgment OFF ON Executing an Averaged Measurement 5 P...

Страница 94: ...ACE Mode Pass judgment Fail judgment The judgment results are also displayed in the event list Turns green when all events are judged as Pass If there is even a single Fail event this turns red In the...

Страница 95: ...eature for Measurement Data Measured waveforms and event analysis data can be saved automatically when averaged measurements are completed To change the save destination folder or file name see Config...

Страница 96: ...displays PDF Measured data is saved as a report file in PDF format When the icon displays SOR PDF Waveform data is saved as waveforms and as a report file simultaneously Saving Waveforms during Avera...

Страница 97: ...d event analysis is automatically executed and the event analysis results are shown on the data display screen For details on averaged measurement see section 4 1 Note If you display waveform data usi...

Страница 98: ...il Summary Display of separate events Displayed by tapping the Detail tab Displays the previous event Displays the next event Soft Key Menu event analysis 4 Tap the soft key menu display button Soft k...

Страница 99: ...ay of the event analysis results Summary List Detail This is the same function as the tab used to switch to display of the analysis results Edit or fix events 2 point markers See page 5 9 Selects the...

Страница 100: ...isplay up to 100 events If you insert an event between two events that are displayed the new event takes on the number that follows the event on the left and all subsequent event numbers are increased...

Страница 101: ...oft key The selected event is deleted The color of the selected event changes The event is deleted virtually When the event is deleted the subsequent event numbers are each decreased by one The delete...

Страница 102: ...t You can control the marker placed on the selected event when the Rotary Knob soft key is set to Cursor 1 Sets marker Places marker at the cursor position 1 Sets marker Places marker at the cursor po...

Страница 103: ...return loss of the selected event Shows the current marker placed when event analysis was executed The selected event is displayed expanded You can edit the splice loss and return loss values by movi...

Страница 104: ...de screen Master mode Registers the current placed event as the master New events are not detected even when the waveform data is updated Only the master event is displayed on the updated waveform dat...

Страница 105: ...n 8 Tap the A soft key Marker A is displayed at the cursor position 9 Turn the rotary knob to move the cursor to the analysis end position 10 Tap the B soft key Marker B is displayed at the cursor pos...

Страница 106: ...point is displayed If the distance reference is specified the distance reference is the measurement reference point If the distance reference is not specified the measurement reference point S is the...

Страница 107: ...nt becomes the E event and a new number is assigned to the former E event Deleting an Event If only the S point and the E event are displayed you cannot delete any events If you delete an event the su...

Страница 108: ...ries depending on the Event Fix Mode as follows When Event Fix Mode Is Set to Append Mode Events detected at positions other than the master event positions are also displayed in the event list Events...

Страница 109: ...ACE Mode and MAP Mode 3 Tap the TRACE MAP button to set the data display screen to MAP mode When the data display screen is in MAP mode the button is TRACE Note If you switch the data display screen t...

Страница 110: ...d in the center position 4 Tap the icon displayed in the center position An event type edit screen appears 5 Select the event type you want to change to Event icon display The event selected in the ev...

Страница 111: ...e loss from the measurement reference point is displayed If the distance reference is specified the distance reference is the measurement reference point If the distance reference is not specified the...

Страница 112: ...MAP button to set the data display screen to TRACE mode When the data display screen is in TRACE mode the button is MAP Note You can operate cursors and markers only in TRACE mode Displaying Waveform...

Страница 113: ...position Note When you turn the rotary knob to the right a cursor appears at the left edge of the data display screen Moving the Cursor 6 Tap the position on the data display screen that you want to...

Страница 114: ...form waveform analysis using markers see the explanation Cursor and marker operation screen If you do not use markers or cursors for about 8 seconds the screen will close automatically Closes the wind...

Страница 115: ...Point Markers 6 Point Markers see Soft Key Menu Markers Marker Analysis Results Splice loss and return loss values are displayed on the screen as marker analysis results For the analysis procedure of...

Страница 116: ...he measurement principle see the explanation 4 Point Monitor Place markers on the waveform data and monitor the loss between S and E and analysis values of up to two events See page 6 10 Preset marker...

Страница 117: ...that the 1 marker is on the measurement start point distance reference side The measured loss value differs depending on the specified approximation method Distance Reference Normally the location whe...

Страница 118: ...n the data screen or turn the rotary knob a cursor appears that you can move Marker n operation When you tap on the data screen or turn the rotary knob marker n appears that you can move Marker E oper...

Страница 119: ...at the cursor position Sets marker Places marker at the cursor position 3 3 Marker Y2 3 Marker Press to set all markers together See 2 Point Markers when the marker mode is Marker Note Set the markers...

Страница 120: ...the cursor position Sets marker Places marker at the cursor position 3 3 Marker Y2 3 Marker See 2 Point Markers when the marker mode is Marker Marker Y1 Marker Y3 Set marker Y1 Places marker Y1 at the...

Страница 121: ...rder shown above starting with the measurement start point the distance reference The measured loss value differs depending on the specified approximation method Set marker E to the correct position T...

Страница 122: ...t A Executes the action on preset B Registering Markers 9 Tap the Action soft key to select Register 10 Tap the A or B soft key The markers applied to the waveform are registered and the marker inform...

Страница 123: ...epending on the approximation method that you have specified This can only be used when Marker Mode is set to Marker Y2 b dB e dB c km f km 3 2 1 Splice loss Return loss b c b c dB km dB km a e f e f...

Страница 124: ...nt 3 At the position of marker 2 the level difference between the approximation line Y1 Y3 and the approximation line Y2 3 is calculated as the splice loss Y2 Y1 Y3 a dB b dB c km e dB f km Approximat...

Страница 125: ...utton is MAP 4 Perform a measurement or load a file to display a waveform on the screen When performing an averaged measurement wait for the measurement to complete When loading a waveform data file u...

Страница 126: ...n horizontally Zooms in vertically Zooms out vertically Zooms out horizontally Cursor and marker operation screen If you do not use markers or cursors for about 8 seconds the screen will close automat...

Страница 127: ...The box indicates where in the entire waveform display area the zoomed waveform display is showing Example of a Full Display Overview display The current data display area is displayed with a frame E...

Страница 128: ...e When the data display screen is in TRACE mode the button is MAP Displaying the Reference Source Waveform 4 Perform a measurement or load a file to display a waveform on the screen When performing an...

Страница 129: ...Taking a Snapshot 8 Tap the Ref Trace soft key The waveform data shown on the data display screen current waveform is captured as a reference trace The reference trace is displayed with a different co...

Страница 130: ...e trace comparison ON Clearing the Reference Trace 12 Tap the Clear Ref soft key The reference trace will be cleared If the differential trace is displayed at this point it will also be cleared Explan...

Страница 131: ...een to TRACE mode When the data display screen is in TRACE mode the button is MAP 4 Perform a measurement or load a file to display a waveform on the screen When performing an averaged measurement wai...

Страница 132: ...sor position Sets the end point Sets the end point marker at the cursor position Specify from markers When the and markers or the n and E markers are already present change these markers to the start...

Страница 133: ...nce point adjustment the start point becomes the reference point B End point Shows marker E The level at the reference point marker S Section analysis values between the start point and end point Sett...

Страница 134: ...nt within the specified section as the reference point The distance return loss total loss and loss per division dB km between the S and E markers are displayed on the screen Marker Auto Setting set f...

Страница 135: ...ent may occur if the light enters the eye French AVERTISSEMENT Lorsque instrument g n re de la lumi re la lumi re est mise travers les ports de source lumineuse Ne pas d brancher les c bles de fibre o...

Страница 136: ...n you tap the OFF button the measurement light turns off Set the wavelength 1310nm 1550nm Set the modulation mode CW 270Hz 1kHz 2kHz Note Check the wavelength and modulation mode settings before turni...

Страница 137: ...t key menu appears Soft key menu Light Source Menu Turns the light on Outputs the measurement light A mark appears on the instrument display to indicate that the light is on Turns the light off Stops...

Страница 138: ...If the LS button is not displayed proceed to step 4 Light source screen Drag to move the light source screen Set the modulation mode CW 270Hz 1kHz 2kHz Turns the light output on and off The output tu...

Страница 139: ...measurement light wavelengths can be generated You can produce continuous light CW or light that has been modulated at the selected frequency modulation mode Model Measurement Light Wavelength Modula...

Страница 140: ...nters the eye French AVERTISSEMENT Lorsque instrument g n re de la lumi re la lumi re est mise travers les ports de source lumineuse Ne pas d brancher les c bles de fibre optique connect s Des l sions...

Страница 141: ...output on and off When you tap the ON button the measurement light turns on The button changes to OFF and a mark appears on the instrument display to indicate that the light is on When you tap the OFF...

Страница 142: ...key menu appears Soft key menu VLS Menu Turns the light on Outputs the measurement light A mark appears on the instrument display to indicate that the light is on Turns the light off Stops the measure...

Страница 143: ...Source Screen Drag to move the visible light source screen Set the modulation mode CW 2Hz Turns the light output on and off The output turns on or off each time you tap the button VLS button Starts t...

Страница 144: ...oft key or set the unit to dB Set the modulation mode CW 270Hz 1kHz 2kHz Performs zero set Bar graph display of the measured value Line indicating the lower threshold Line indicating the upper thresho...

Страница 145: ...number OFF a b 2 a c 3 a d 4 a e 5 a f 6 a g 7 a h 8 Set the measurement interval 500ms 1s 2s 5s 10s Set the number of times 10 to 36000 Set the average count 1 10 50 100 Set the offset 9 900 to 9 900...

Страница 146: ...en For the position of the OPM port see Component Names and Functions in the Getting Started Guide IM AQ1210 02EN Soft Key Menu optical power meter 9 Tap the soft key menu display button An optical po...

Страница 147: ...sured value The display changes to show values relative to the reference value dB If the absolute measurement value dBm after executing DREF is the same as a reference value 0 dB is displayed The meas...

Страница 148: ...the measurement conditions Saved data information You can save up to three sets of data in the save area of the specified core Note You can save the optical power meter measurement results by followi...

Страница 149: ...t of Optical Power Values logging 10 Tap the Logging soft key A logging measurement screen appears Optical power value vertical axis Cursors C1 and C2 Thresholds Measurement count The number of times...

Страница 150: ...pears for the auxiliary function of the logging screen Clears the cursor Set the threshold display OFF ON Auto zoom Automatically zooms in on the graph at the cursor position Zoom Initialize Resets th...

Страница 151: ...cal power meter screen Drag to move the optical power meter screen Set the modulation mode CW 270 Hz 1 kHz 2 kHz OPM button Starts the optical power meter screen Returns the waveform display area to i...

Страница 152: ...is preceded by a prefix m 10 3 10 6 n 10 9 or p 10 12 The following relation holds between the absolute display unit dBm and W PdBm 10 log Pw 103 PdBm optical power unit dBm Pw optical power unit W Re...

Страница 153: ...When set to off the maximum and minimum values are reset Offset You can add a specified value the offset value to measured input optical power values and display the results as measured values The ra...

Страница 154: ...lected the corresponding core is not measured By specifying Skip on cores that do not need to be measured you can save time Executing a Save Operation When you execute a save operation up to three set...

Страница 155: ...STOP soft key logging stops The Logging STOP soft key changes to Logging START and the file save screen appears The logging results can be saved in csv format Interval Set the interval for logging the...

Страница 156: ...e down arrow key zooms out Zoom factors are 1 2 5 10 20 and 50 Left and right arrow keys Zooms in or out horizontally The right arrow key zooms in the left arrow key zooms out Zoom factors are 1 2 5 1...

Страница 157: ...Power Meter Screen Set the transmission direction ONU OLT wavelength 1310 nm OLT ONU wavelength 1490 nm and 1550 nm When you set the transmission direction the wavelength is automatically set Measured...

Страница 158: ...e upper and lower limits for each wavelength of the PON system OPM Setup tab Executing the Zero Set 4 After step 3 close the optical power meter screen The screen returns to the PON power meter screen...

Страница 159: ...wer meter screen For the position of the OPM port see Component Names and Functions in the Getting Started Guide IM AQ1210 02EN Soft Key Menu PON power meter 8 Tap the soft key menu display button An...

Страница 160: ...9 4 Measured data measurement conditions and measured values Shows the optical power value currently being measured and the measurement conditions Saved data information You can save up to three sets...

Страница 161: ...ne termination on the subscriber side to the OLT Optical Line Termination optical line termination on the station side Measures the optical power of the signal whose wavelength is 1310 nm for data 2 M...

Страница 162: ...the Results of Optical Power Measurements for Multi Fiber Cables Skip When Skip is selected the corresponding core is not measured By specifying Skip on cores that do not need to be measured you can s...

Страница 163: ...ker to display the POWER CHECKER screen MENU screen Power Checker Screen Set the wavelength SM 1310nm SM 1490nm SM 1550nm SM 1625nm SM 1650nm Measured value Set the unit dBm W Set the reference value...

Страница 164: ...procedure Measuring the Optical Power 4 Set the wavelength 5 Connect an optical fiber cable to the OTDR port PORT1 The measured value is displayed on the power meter screen For the position of the OT...

Страница 165: ...reference value is set to the current measured value and the measurement display changes to dB units The measurement display indicates a value relative to the reference value Measured value The displa...

Страница 166: ...creen Drag to move the power checker screen OPC button Starts the power checker screen Returns the waveform display area to its original condition The example of the OTDR screen shows the waveform dis...

Страница 167: ...t changes to dB When you tap the DREF soft key or set the unit to dB the Reference box appears on the power meter screen You can set the reference value in the Reference box The range is 80 dBm to 40...

Страница 168: ...p Fiber Inspec Probe to display the FIBER INSPEC PROBE screen MENU screen Fiber Inspection Probe Screen Shows the fiber end face image The model name of the probe is appeared on the display when the p...

Страница 169: ...Select the judgment reference When the fiber type is SM UPC SPC APC When the fiber type is MM default Set the defect judgment threshold 0 to 100 You can set the value when the defect judgment executio...

Страница 170: ...n probe 5 Tap the soft key menu display button A soft key menu for the fiber inspection probe appears Soft key menu Fiber Inspection Probe Menu Displays the image in real time Holds the image display...

Страница 171: ...esults detail Starts saving Press this soft key to save a still image and a judgment result file For the procedure to set the file save destination file type and file name see section 9 4 Optical fibe...

Страница 172: ...een This is a video screen Displays the still image Displays the still image over the entire data display screen FIP button Starts the fiber inspection probe screen Returns the waveform display area t...

Страница 173: ...or off If you connect a USB fiber inspection probe while the instrument is on the instrument will automatically identify the probe For cautions regarding connecting the probe see Note in section 9 1 F...

Страница 174: ...e Scratch or Defect Judgment Threshold Values Set the number of scratches and defects that will be judged as fail The range is 0 to 100 If the judgment execution check box explained above is cleared N...

Страница 175: ...is mistakenly emitted from these ports enters the eye French AVERTISSEMENT Lorsque instrument g n re de la lumi re la lumi re est mise travers les ports de source lumineuse Ne pas d brancher les c bl...

Страница 176: ...50 nm 1300 nm 1625nm 1650nm Set the fiber in use alarm OFF ON Set auto save MEASURE tab Initializes the measurement settings Set the connection check OFF ON Set the distance range AQ1210A AQ1210E AQ12...

Страница 177: ...el Executes initialization Press to reset the analysis settings to their factory defaults ANALYSIS tab Set macro bending OFF ON Set the pass fail judgment OFF ON Set the approximation method event TPA...

Страница 178: ...m display that is controlled with the arrow keys dB decimal places Select the cumulate loss type Type1 Type2 Type3 Select the calculation method for cumulate loss Turns approximated line display on or...

Страница 179: ...setup screen that appears when you select a setting and the descriptions of the settings see section 2 4 Executing an Averaged Measurement 5 Tap Wavelength to set the wavelength 6 Press AVG A measurem...

Страница 180: ...t Analysis in section 1 4 View and edit waveforms in trace mode Viewing and Editing Waveforms in Trace Mode 7 Tap the TRACE MAP button to set the data display screen to TRACE mode A waveform view edit...

Страница 181: ...appears for operating the core selection screen of multi fiber measurement You cannot operate in MAP mode Soft key menu Waveform View Edit Menu Edit the composite waveform display Edit the display ar...

Страница 182: ...en and move them Moves the target to the left Moves the line selected in the target section to the left Returns the lines to their original positions Moves the target to the right Moves the line selec...

Страница 183: ...light that is mistakenly emitted from these ports enters the eye French AVERTISSEMENT Lorsque instrument g n re de la lumi re la lumi re est mise travers les ports de source lumineuse Ne pas d branch...

Страница 184: ...pears in half of the cell If the measurement results of two waveforms are saved the color appears over the entire cell Waveform data Test data of the fiber end face Measured data of the optical power...

Страница 185: ...See page 8 26 Set the wavelength for the real time measurement or averaged measurement 1310nm 1550nm 1625nm 1650nm You can select from the wavelengths set in the setup measurement Measure conditions...

Страница 186: ...he same label can be entered for all cores See section 2 4 Project information You can view the project settings and measurement settings in a table See page 8 26 Set the file name type WL No No WL Pr...

Страница 187: ...F ON Set the average times when the average unit is set to Times AUTO 2 10 2 11 2 12 2 13 2 14 2 15 2 16 2 17 2 18 2 19 2 20 Set the distance range AQ1210A AQ1210E AQ1210D 1310 1550 nm AUTO 200m 500m...

Страница 188: ...o reset the analysis settings to their factory defaults ANALYSIS tab Set macro bending OFF ON Set the pass fail judgment OFF ON See page 2 8 Set the approximation method event TPA LSA Set the splitter...

Страница 189: ...hat is controlled with the arrow keys dB decimal places Select the cumulate loss type Type1 Type2 Type3 Select the calculation method for cumulate loss Turns approximated line display on or off Displa...

Страница 190: ...set to on in the setup measurement Measure conditions the analysis results are displayed when the measurement is complete Note If Wavelength on the multi fiber measurement screen is set to All measure...

Страница 191: ...en a message appears for confirming the saving of the measured results If necessary save the measured results Soft Key Menu averaged measurement 7 Tap the soft key menu display button A soft key menu...

Страница 192: ...Progress Returns to the multi fiber measurement screen core selection screen Marker operation See section 6 1 Event analysis control See section 5 1 Section analysis control See section 6 4 8 2 Measur...

Страница 193: ...p of the screen Laser on indication The OTDR port emitting light Real time measurement in progress indication Core number being measured OTDR waveform screen Cannot be controlled on the OTDR waveform...

Страница 194: ...label setting of the project setup screen the same label characters can be entered at once for all cores Soft Key Menu real time measurement 7 Tap the soft key menu display button A soft key menu app...

Страница 195: ...Is Not in Progress Returns to the multi fiber measurement screen core selection screen Marker operation See section 6 1 Event analysis control See section 5 1 Section analysis control See section 6 4...

Страница 196: ...saved Performing Setup 7 Press SETUP to display the optical power meter setup screen 8 Tap the Power Meter tab to display the POWER METER screen For details on how to set up the optical power meter se...

Страница 197: ...ft Key Menu optical power meter 13 Tap the soft key menu display button An optical power meter soft key menu appears Soft key menu Optical Power Meter Menu Sets the reference to the currently displaye...

Страница 198: ...o A menu appears for executing a measurement Example of measuring core number 44 Fiber inspection probe 7 Tap the FIP menu to display the FIBER INSPEC PROBE screen For details on how to use the fiber...

Страница 199: ...Menu Displays the image in real time Holds the image display HOLD Starts saving Press this soft key to save the still image to a BMP file For the file save destination see section 9 4 For the procedu...

Страница 200: ...Fail Judgment Is Set to OFF Check marks appear next to the wavelengths for which measured data is saved Check marks Preview waveform When Pass Fail Judgment Is Set to ON Judgment result marks appear...

Страница 201: ...he settings of the current project and creates a new project Takeover current setting Duplicates the settings of the current project to create a new project Cancel Cancels the creation of a new projec...

Страница 202: ...TDR Emulation Software can be loaded into the instrument Note On the instrument you can only create projects whose measurement conditions of all cores are the same Multi Fiber Project can be used to c...

Страница 203: ...light that is mistakenly emitted from these ports enters the eye French AVERTISSEMENT Lorsque instrument g n re de la lumi re la lumi re est mise travers les ports de source lumineuse Ne pas d branch...

Страница 204: ...Loss Test tab Set the number of cores or number of tapes up to 100 cores when the tape number type is Off up to 50 tapes when the type is a b 2 up to 12 tapes when the type is a h 8 Set the starting...

Страница 205: ...Configure the power meter display screen Sets the reference to the currently displayed measured value See page 7 11 Holds the measurement values See page 7 11 Saves data A screen appears for saving t...

Страница 206: ...lue of wavelength 1 Measured value of wavelength 2 Line indicating the lower threshold Line indicating the upper threshold Performs zero set Set the unit dB dBm W See section 7 3 Executing the Zero Se...

Страница 207: ...art a loss test the instrument generates the specified wavelengths in order The current wavelength appears here Starts a loss test OFF ON A loss test starts when this is set to ON The loss test stops...

Страница 208: ...anges to Loss Test STOP Then on the power meter side of the instrument the optical power value of the wavelength is displayed on the screen On the light source side of the instrument tap the Loss Test...

Страница 209: ...pback display screen you can operate the light source screen and the optical power meter screen For descriptions of the display screens see page 8 30 optical power meter and page 8 31 light source Opt...

Страница 210: ...er measurement section and enables you to obtain more accurate absolute optical power values Adjusting the Optical Power 8 Before starting the loss test directly connect the OTDR port PORT1 of the ins...

Страница 211: ...hanges to Loss Test STOP Then the optical power value of the wavelength is displayed on the instrument screen Tap the Loss Test STOP soft key to end the loss test Note When you start a loss test the o...

Страница 212: ...to execute the optical power adjustment When adjustment ends normally the Instrument returns to the previous screen While adjusting the word Execute on the menu changes to Abort All soft keys other th...

Страница 213: ...ght that is mistakenly emitted from these ports enters the eye French AVERTISSEMENT Lorsque instrument g n re de la lumi re la lumi re est mise travers les ports de source lumineuse Ne pas d brancher...

Страница 214: ...slave side the project conditions sent from the master side are received and loaded into the instrument Setting Indicates that the master and slave are being selected or the project information is bei...

Страница 215: ...s are being received the Receive Project from Master soft key changes to Stop When reception is complete the Stop soft key changes back to Receive Project from Master and the loss test screen appears...

Страница 216: ...9900 Set the number of cores or tapes 1 to 2000 Set the tape number OFF a b 2 a c 3 a d 4 a e 5 a f 6 a g 7 a h 8 Set wavelength 1 GI 850nm GI 1300nm SM 1310nm SM 1550nm SM 1625nm SM 1650nm Set the o...

Страница 217: ...instrument to the OPM port with a short optical fiber cable AQ1210A AQ1215A AQ1210D AQ1210E AQ1215E AQ1215F AQ1216F Light source Optical power meter Short optical fiber cable OTDR port PORT1 OPM port...

Страница 218: ...ster Create a project Slave Optical fiber cable for signal transmission OTDR port PORT1 OPM port Transfer the project data 9 On the slave side tap the Receive Project from Master soft key The soft key...

Страница 219: ...ort Transfer the results of the multicore loss test Optical fiber cable under multicore loss test OTDR port PORT1 OPM port Light being measured Slave Master 12 Tap the core number you want to perform...

Страница 220: ...test before the loss tests have been completed on all cores If you want to resume the loss test using the same shared project the test can be resumed while retaining the data of the completed tests Da...

Страница 221: ...istakenly emitted from these ports enters the eye French AVERTISSEMENT Lorsque instrument g n re de la lumi re la lumi re est mise travers les ports de source lumineuse Ne pas d brancher les c bles de...

Страница 222: ...display a waveform obtained by subtracting the values of one waveform from those of another waveform selected as the current trace Multi Trace Analysis 3 Tap the Multi Trace item to display the Multi...

Страница 223: ...aveform to place markers and cursors on or move vertically Moves the trace up Moves the display position of the current trace up Moves the trace down Moves the display position of the current trace do...

Страница 224: ...itches the display of the marker analysis values Current Trace 1 2 2 3 Splice loss Return loss Note The current waveform display before switching to the multi trace analysis screen is read into the tr...

Страница 225: ...Trace screen Overview display See section 6 2 Trace1 waveform Trace2 waveform List of events Displayed by tapping the List tab For details on events see the explanation in section 5 1 Display of separ...

Страница 226: ...creen When Merged is selected the event information is combined and displayed Combines the waveforms of each direction Combined with the waveform viewed from the other direction The display direction...

Страница 227: ...waveform TraceA waveform display TraceB waveform display 2 way trace waveform display displays the event number of TraceA of the current trace 2 way trace waveform display Combines and displays the e...

Страница 228: ...ances to the marker positions on the waveform Markers for monitoring loss values Differential trace Marker information of the trace waveform specified as the current trace Switches the loss value disp...

Страница 229: ...eforms Clears the waveforms of Trace1 to Trace2 When tapped a file operation screen appears and you can load waveforms See section 9 4 4 Point Monitor This is an example of placing 4 Point Monitor mar...

Страница 230: ...ormation are those of the current trace 2 way Trace Analysis Loading Waveform Data The type of file that you can load is SOR Traces that meet the following conditions can be combined Event lists are a...

Страница 231: ...traces that are used for taking the difference Cursors are displayed on the current trace Distance and splice loss that are displayed in the marker information are those of the current trace Different...

Страница 232: ...se a portable USB memory device for the USB storage device Connect it directly to a type A USB port of the instrument Hot plugging is supported you can connect or disconnect the USB device at any time...

Страница 233: ...ge For the procedure see section 10 1 5 On your PC start Explorer or another browser Instrument s internal memory Select Removable Disk Note The ends of the USB cable supplied with the product are bot...

Страница 234: ...LE Icon You can display the data management buttons using the FILE icon shown on the data display screen of the OTDR power meter and the like This FILE icon can be used with all features OTDR and othe...

Страница 235: ...e Available Unavailable dimmed Available PON power meter Available Available Unavailable dimmed Available Power checker Unavailable dimmed Unavailable dimmed Unavailable dimmed Available Light source...

Страница 236: ...ain d tre enregistr es ou charg es partir d une m moire interne Vous risqueriez d endommager le support de stockage m moire USB ou m moire interne ou les donn es qu il contient Procedure Saving Data e...

Страница 237: ...file operation see section 9 6 Saving the System Setup Data CFG format 1 Press MENU to display the MENU screen 2 On the data management button window tap SAVE A file list screen for saving data appear...

Страница 238: ...lays the contents of the next higher level folder File path List of files and folders Loads data Sorts the file list Note The instrument can load SOR files waveform data and CFG files configuration da...

Страница 239: ...event Event data is saved to a file in CSV format SMP Adapt Trace data measured from an optical pulse including the measurement conditions is saved to a file in SMP format The screen image immediately...

Страница 240: ...xtension Description SOR File in a format that conforms to Telcordia SR 4731 or Bellcore GR 196 CORE must be a data file with measurement conditions that can be set on the instrument SMP Adapt Trace d...

Страница 241: ...s l alimentation lectrique lorsque l indicateur d acc s la m moire USB clignote ou lorsque les donn es sont en train d tre enregistr es ou charg es partir d une m moire interne Vous risqueriez d endom...

Страница 242: ...lders Exports a report file Name of the file to be saved See Setting the File Name in section 2 4 Sorts the file list Note The waveform shown on the data display screen is exported to a PDF report fil...

Страница 243: ...ta management buttons 4 Tap FILE to display the file list screen File Operations Data management buttons 5 Select the waveform data file to export to a report file For instructions on how to select mu...

Страница 244: ...view PDF report files you need Adobe Reader 5 0 or later Setting the Report File Format See Setting the Report Format or Setting the File Report Format in section 2 4 Explanation This section explain...

Страница 245: ...limentation lectrique lorsque l indicateur d acc s la m moire USB clignote ou lorsque les donn es sont en train d tre enregistr es ou charg es partir d une m moire interne Vous risqueriez d endommager...

Страница 246: ...ting files and folders to be manipulated rename copy delete files and folders Sorts the file list Renaming Copying and Deleting Files and Folders 3 Tap the file or folder you want to manipulate A scre...

Страница 247: ...y destination file list Copy destination example Under the Sample folder Returns to the previous screen Selecting Multiple Files and Folders 3 Tap the icon A screen for selecting files and folders to...

Страница 248: ...e you want to load A screen with file operation icons shows a load file icon Loads files The selected files will be loaded Select the file The background color changes when selected Returns to the pre...

Страница 249: ...ted from Set the language See Selecting the Language to Display in the Getting Started Guide IM AQ1210 02EN Set the power save mode See section 10 2 Set the start screen Top Menu Start Menu to Last Fu...

Страница 250: ...OFF Beep sound is off Start Up Window Set the first screen that appears when you start the instrument Top Menu The MENU screen appears Start Menu to Last Function The screen of the feature that was ex...

Страница 251: ...ark environment The battery lasts longer than the Normal setting OFF The LCD backlight is turned off The battery lasts longer even longer than the Power Save setting When you press any key the backlig...

Страница 252: ...on dongle USB Type A 2 ports LAN cable Note Please use your own communication dongle For the recommended communication dongles that can be used with this instrument contact your nearest YOKOGAWA deale...

Страница 253: ...rk Valid Invalid Set the timeout period Infinite 0 1 to 3600 Set the DHCP OFF ON Set the TCP IP parameters Applies the network settings Set the Common Setup 5 Tap Common Setup to display the Common Se...

Страница 254: ...feature You do not need to enter a password when the user name is set to anonymous Time Out If there is no access to the instrument within the specified period the network connection between the inst...

Страница 255: ...dentification Consult your network administrator for the netmask address If DHCP is available and you set DHCP above to ON the IP address is assigned automatically Gateway Set the IP address of the de...

Страница 256: ...dongle For the recommended communication dongles that can be used with this instrument contact your nearest YOKOGAWA dealer Network Setup Screen 2 Press MENU to display the MENU screen 3 Press SETUP t...

Страница 257: ...ructions on how to input characters see the Getting Started Guide IM AQ1210 02EN Set the authentication password A character input screen appears For instructions on how to input characters see the Ge...

Страница 258: ...tions on how to input characters see the Getting Started Guide IM AQ1210 02EN Connect to Access Point Connect to an access point Set the network address A numeric keypad appears Use a fixed ID name SS...

Страница 259: ...ess operation does not end and the icon remains displayed In addition even if you stop accessing from the wireless terminal the wireless LAN operation does not end so you can access from the wireless...

Страница 260: ...cessing from the wireless terminal the wireless LAN operation does not end so you can access from the wireless terminal again Operation on the Wireless Terminal 6 Search for the access point from the...

Страница 261: ...els waveform data transfer Access point information An icon indicating that the wireless LAN is in use If you execute the Stop WLAN function the wireless LAN operation ends and the icon disappears If...

Страница 262: ...y a single wireless terminal can access the instrument at any given time Multiple wireless terminals cannot access the instrument simultaneously Operation on the Wireless Terminal Web Browser 6 Search...

Страница 263: ...selection Displays a file dialog box Screen refresh interval shows a setup screen Corresponds to the AQ1210 panel keys Note If you set the user name in the network settings to anonymous enter anonymo...

Страница 264: ...he communication with the wireless terminal None An encryption key is not used WPA2_PSK A WPA2 PSK encryption key is used Broadcast Channel Beacon Signal Specify the channel on which to receive beacon...

Страница 265: ...kogawa co jp tm F SOFT Waveform data can also be transferred by remotely controlling the instrument See Displaying the File Dialog Box on the previous page Remote Controlling the OTDR Operation on the...

Страница 266: ...tapping by clicking the mouse Dragging and pinching are not possible Downloading Files Only files that are 1400 KB or less in size can be downloaded The download button does not appear for files that...

Страница 267: ...Expiration Date Setup screen Set the expiration date and time When you set the expiration date and time the number of remaining days until expiration is displayed Set the action after expiration Lock...

Страница 268: ...d in Entering Text in the Getting Started Guide IM AQ1210 02EN Enabling the Expiration Date Settings 6 Continuing from step 3 tap Expiration Release Code to display a number input screen 7 Enter a rel...

Страница 269: ...the Locked State When the expiration date is reached a screen indicating that the instrument is locked is displayed at startup 1 Tap Expiration Release Code to display a number input screen 2 Enter a...

Страница 270: ...umber when the instrument starts If you do not enter the correct number the instrument will not start Note that if the expiration date arrives while you are using the instrument startup completed cond...

Страница 271: ...y if there is no user activity for the specified time Check the settings 11 2 The screen is dark The screen will appear dark when the LCD Brightness setting is set to Power save Check the settings 11...

Страница 272: ...3 Data invalid 500 The measurement condition cannot be replicated Please modify the measurement setup where is shown 501 Not executable during measurement Please stop the measurement and execute again...

Страница 273: ...rmed Please check media is correctly installed When using USB memory Please try the other media or check the USB memory is formatted with FAT 653 Same file name or folder name exists Please change the...

Страница 274: ...condition that cannot be measured with this instrument Pulse Width 732 ID No is not selected in File Name Please include ID No in File Name if using direct save Select ID 819 ROM Test Error occurred 8...

Страница 275: ...The instrument will be powered off in 10 sec Please make sure of using the dedicated the power adapter 944 The voltage of a module is too high The instrument may be damaged if it is kept using in thi...

Страница 276: ...display the MENU screen The data management buttons appear 2 Tap the icon to display the product information screen MENU screen Data management buttons Product Information Screen Saves the hardware in...

Страница 277: ...length Version Information The firmware version and update date information are displayed Screen Capture The screen image of the product information can be saved in internal memory Folder name Product...

Страница 278: ...g the SETUP key on screens other than the MENU screen such as those of the OTDR feature and power meter feature Depending on the feature you may need to tap the System tab after displaying the setup s...

Страница 279: ...Displays the contents of the next higher level folder File path Sorts the file list Explanation Firmware update files are YMC files e g AQ1210 YMC If necessary you can change the file name but on the...

Страница 280: ...reen by pressing the SETUP key on screens other than the MENU screen such as those of the OTDR feature and power meter feature Depending on the feature you may need to tap the System tab after display...

Страница 281: ...internal memory will also be deleted Initializes settings The setup data will be initialized Note The USERS_MANUAL folder in the USER folder and the PDF user s manuals in the USERS_MANUAL folder will...

Страница 282: ...n the feature you may need to tap the System tab after displaying the setup screen with the SETUP key The settings are applied globally regardless of which screen system setup is executed from 3 Tap O...

Страница 283: ...vec instrument sous tension peut permettre la lumi re mise de p n trer accidentellement dans les yeux Ceci peut provoquer des l sions oculaires ou une d ficience visuelle ATTENTION Si des corps trange...

Страница 284: ...he light emitter turn off the instrument to prevent the light from accidentally being emitted from the light source port Cleaning the light emitter with the instrument turned on may cause the emitted...

Страница 285: ...0 02EN 4 Use a lens cleaner to clean the optical fiber end face 5 Insert the lens cleaner slowly into the optical adapter s optical connector and turn the lens cleaner to clean the inner walls Lens cl...

Страница 286: ...nstrument 1 Wrap the instrument in a thick plastic sheet or the like to prevent dust from entering the inside of the instrument 2 Apply cushioning material to the LCD to protect it 3 Prepare a box wit...

Страница 287: ...product includes open source software For the open source software licenses see TermsAndConditions_OpenSourceSoftware pdf in the USER DATA USERS_MANUAL folder of this instrument For instructions on ho...

Страница 288: ...eport Setup 2 29 files loading 9 17 files saving and loading 1 28 FINE 6 13 FIP utility 7 38 Firmware Update 11 8 Fresnel reflection 1 8 Index Numeric Page 2 Point Markers 6 12 2 Point Markers line 6...

Страница 289: ...1 24 Multi Core Loss Test 1 4 multicore loss test menu 8 41 multi core optical fiber 1 21 multi fiber 1 4 Multi Trace 8 48 multi trace analysis 1 25 multi trace analysis menu 8 49 multi wavelength mea...

Страница 290: ...m 1 5 T Page tape number 1 21 timeout 1 30 Time Out network 10 6 To be set by Marker 6 22 total loss 2 21 total return loss 2 21 TPA 2 12 TRACE mode 1 2 TRACE mode analysis 1 10 TRACE mode display 1 7...

Отзывы: