Executing a Loss Test
10.
Loop back the optical fiber cable connected to the OTDR port (PORT1) of the instrument at the
far end of the cable installation, and connect the other end of the optical fiber cable to the OPM
port of the instrument.
AQ1210A, AQ1215A,
AQ1210D, AQ1210E,
AQ1215E, AQ1215F,
AQ1216F
Light source
Optical power meter
Optical fiber cable
under loss test
OTDR port (PORT1)
OPM port
Loopback
11.
Tap the
Loss Test START
soft key. The loss test starts, and the soft key display changes to
Loss Test STOP. Then, the optical power value of the wavelength is displayed on the instrument
screen.
Tap the Loss Test STOP soft key to end the loss test.
Note
When you start a loss test, the optical power values of wavelength 1 and wavelength 2 (if the wavelength is
set) are measured once. Then, the soft key display changes to Loss Test STOP, which you can use to end
the loss test. In addition, you can abort the loss test by tapping the loss test start ON/OFF button on the light
source display screen.
Saving Data
12.
Tap the
Save Data
soft key. A screen appears for saving the results of optical power
measurements for multi-fiber cables. For details, “Saving the Results of Optical Power
Measurements for Multi-Fiber Cables” (page 7-12) in section 7.3.
8.3 Performing an Auto Loss Test (Auto Loss Test)