8-4
IM AQ1210-01EN
Display (OTDR) Conditions
4.
Tap the
OTDR
tab to display the following screen.
1 2
3
Set the calculation method for total loss (Cumulate loss, Loss between S and E).
Cumulate: Integrated value of events
Between two points: Loss between S and E
Marker information display (OFF, ON)
Display the loss and the distance from the measurement reference point to each marker on the waveform
display area.
Ghost cursor display on/off
Checks secondary reflections
Set the marker type (Marker, Line).
Marker:
Line:
OTDR tab
Set the zoom method (Natural, Legacy).
Select the zoom behavior of the waveform display that is controlled
with the arrow keys.
dB decimal places(**.*, **.**, **.***)
Select the cumulate loss type (Type1, Type2, Type3)
Select the calculation method for cumulate loss.
Turns approximated line display on or off
Displays an approximated line for waveform
events.
Select the measurement
distance unit
This menu is not available
on -HJ suffix code.
Select the reflection term
(Return Loss, Reflection
level)
This menu is available on
-HJ suffix code.
Calculation method for
total return loss (Include
END, Exclude END).
END point loss display
(ON, OFF)
Distance decimal places
(**.***, **.****, **.*****)
Cursor dB value display
(ON, OFF)
Locks screen operations
Screen operations can be
locked after a given time
elapses.
Select the Cursor
display format
(CROSS(+), LINE(¦ ))