11-4
IM AQ1210-01EN
Code
Message
700
Failed to open the file.
701
Failed to close the file.
702
Failed to read the file.
703
Irregular file format.
704
Failed to write the file.
705
Cannot be saved. No trace data.
708
Cannot be saved. This data is not taken with this instrument.
710
File cannot be retrieved. Invalid wavelength
711
File cannot be retrieved. Invalid distance range
712
File cannot be retrieved. Invalid pulse width
713
File cannot be retrieved. Invalid sampling points
714
File cannot be retrieved. Distance range exceeds 400km.
715
File cannot be retrieved. Sampling intervals exceed 64m.
716
File cannot be retrieved. Actual averaging times or duration is not set.
717
Duplicate file name.
718
File is damaged. Check the file.
719
File name too long. Maximum length is 60 letters.
720
File is now being accessed. Execute after access is released.
721
Cannot load this file. Invalid file format or this firmware version is old.
722
Recall Setup File cannot load the measurement condition which sampling interval has been changed to shorter
than the standard sampling interval at Meas. Range Change function.
723
This file cannot be supported by the firmware with this product. Please update into the latest firmware to read this
file.
724
The model name of the instrument and model name in the file is different. This file cannot be read with this
instrument.
725
Can't use the following characters. \ / : ; * ? " < > | % . ,
726
This file cannot be read with this instrument. This file that tries to be loaded includes the measurement condition
that cannot be measured with this instrument.
727
This file cannot be read with this instrument. This file that tries to be loaded includes the measurement condition
that cannot be measured with this instrument. - Distance Range
728
This file cannot be read with this instrument. This file that tries to be loaded includes the measurement condition
that cannot be measured with this instrument. - Pulse Width
732
ID No. is not selected in File Name. Please include ID No. in File Name if using direct save(Select ID).
819
ROM Test Error occurred.
820
RAM Test Error occurred.
821
System File Test Error occurred.
822
Battery Test Error occurred.
823
Temperature Test Error occurred.
824
Power Supply Test Error occurred.
825
WLAN Test Error occurred.
854
In USB Storage mode, all keys are locked. Please disconnect the USB Cable.
902
Battery is low. Please power it off, and charge the battery or replace the battery. Or, please use the power supply.
903
Backup battery is low.
904
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
906
Battery is low. The instrument will be powered off in 10 sec.
909
The temperature inside the instrument is too high. The instrument may be damaged if used in this condition.
The instrument will be powered off in 10 sec. Please turn off power to allow instrument to cool down.
919
The operation of the battery is not confirmed normally. After stopping the measurement, turn off the power.
Please check the connection of battery.
922
Incorrect date and time setting. Set the correct date and time.
925
Please use AC adapter.
933
Setting conflict.
934
Communication not allowed during measurement
935
USB port over current error.
936
Invalid license key. Option was not installed.
937
The temperature inside the instrument is too high. The instrument may be damaged if used in this condition.
The instrument will be powered off in 10 sec. Please turn off power to allow instrument to cool down.
938
The temperature inside the instrument is too low. The instrument may be damaged if it is used in this condition.
The instrument will be powered off in 10 sec. Please turn off power until internal temperature is increased.
939
The temperature inside the instrument is too high. The instrument may be damaged if used in this condition.
The instrument will be powered off in 10 sec. Please turn off power to allow instrument to cool down.
11.2 Error Message Display