2-20
IM AQ1210-01EN
Marker Information
You can display the distance from the measurement reference point to each marker on the waveform
display area.
OFF: Marker information is not displayed.
ON: Marker information is displayed.
Marker information
Reflection Display (when the suffix code is -HJ)
You can set the reflection display at the calculation result display area of measurement data to either
of the following settings.
Return loss: The ratio of the incident optical power level and the reflected optical power level is
displayed.
Reflection level: The reflected optical power level is displayed.
Reflection display
Cursor dB Value
You can display not only the distance at the cursor position, but also the dB value.
OFF: Only the distance at the cursor position is displayed.
ON: he distance and the dB value at the cursor position are displayed.
dB value at the cursor position
2.3 Display (OTDR) Conditions