8-8
IM AQ1210-01EN
Editing the Composite Waveform Display
4.
Tap the
Waveform Edit
soft key. A soft key menu appears for operating the composite
waveform display.
Target sections (Line1, Line2, Line3)
The number of lines varies depending on the number of measured pulse widths. In this
screen example, the number of target segment lines is two because the number of
pulse widths is three. You can also tap the lines directly on the screen and move them.
You can also tap the lines directly
on the screen and move them.
Moves the target to the left
Moves the line selected in the
target section to the left
Returns the lines to
their original positions
Moves the target to the right
Moves the line selected in the
target section to the right
Other waveform display
Displays all waveforms measured
at each pulse width
20 ns section
(example)
500 ns section
(example)
20 µs section
(example)
Pulse width of the
measured Adapt Trace
waveform
Explanation
Event Icon Display
The following events are displayed using icons.
Splice loss
Return loss
Bending loss Splitter
End point
(Fresnel
reflection)
Start point
(Near-end
reflection)
8.1 Displaying a Map of the Line Configuration and Events (OTDR Smart Mapper)