5-8
IM AQ1200-01EN
Marker Mode
Marker
Move the cursor to the location that you want to measure, and set a marker.When you are
measuring return loss and splice loss, they are calculated automatically when you set all the
markers that are necessary for the measurement method that you are using.
Line
After you select the line marker that you want to operate from a set of multiple line markers, move
the selected line marker directly. The value of the line marker’s section is calculated, and the return
loss and splice loss are measured. Each value is calculated in real time as you move the line
marker.
Marker
Line
Note
Line is often used outside of Japan. Marker is often used within Japan. It is used on existing YOKOGAWA
models.
Displaying the Cursor dB Values
You can display not only the distance at the cursor position, but also the dB value.
Off: Only the distance at the cursor position is displayed.
On: The distance and the dB value at the cursor position are displayed.
dB value
Displaying the Scale
You can display the scale values on the vertical and horizontal axes.
Off: The scale values are not displayed.
On: The scale values are displayed.
Scale display on
Scale display off
Scale value
5.5 Setting the Waveform Information Display