12-2
IM AQ1200-01EN
Explanation
This feature is available on models with the /SLT or /HLT option.
You can create a new project.
Creating a New Project
Only the master can create a new project. You can specify the following project information.
Project Name
You can set the name using up to 30 characters.
Starting Core Number
You can set the starting core number to a value within the following range.
1 to 9900
Tape Number Type
Off, a-b(2), a-c(3), a-d(4), a-e(5), a-f(6), a-g(7), a-h(8)
Number of Fibers or Tapes
You can set the number of fibers or tapes to a value within one of the following ranges.
Tape Number Type
Number of Fibers or Tapes
Off
10 to 100 fibers
a-b(2)
10 to 50 tapes
a-c(3)
10 to 33 tapes
a-d(4)
10 to 25 tapes
a-e(5)
10 to 20 tapes
a-f(6)
10 to 16 tapes
a-g(7)
10 to 14 tapes
a-h(8)
10 to 12 tapes
Measurement Light Wavelength
The following measurement light wavelengths can be generated.
Model
Measurement Light Wavelength
AQ1200A
SM 1310 nm, SM 1550 nm
The light for a single mode (SM) optical fiber is emitted from optical port 2. Firmly connect the
optical fiber to the port that the light will be transmitted from.
Offset
See “Explanation” in section 9.2.
Skipping
If you specify Skip for a core number that you don’t need to measure, its data will not be saved. By
setting which cores to skip beforehand, you can avoid accidentally saving unnecessary data.
Saving Project Information
You can save project information to a file. For the procedure for saving project information to a file, see
section 16.2. You can load saved files to both the master and the slave (see section 12.2).
12.1 Creating New Projects