11-4
IM AQ1200-01EN
11.2 Performing an Auto Loss Test
Configuring the Light Source and Executing an Auto Loss Test
Light Source Screen
1.
Using the
rotary knob
and
ENTER
, select
Auto Loss Test
.
2.
Press the
Function Select
soft key and then the
Light Source
soft key to display the following
screen.
When you start a loss test, the AQ1200 generates the
specified wavelengths in order. The current
wavelength appears here.
Optical power adjustment (see section 11.1)
Set the wavelength (SM 1310nm, SM 1550nm).
Executing an Auto Loss Test
WARNING
• While the AQ1200 is producing light, do not remove the optical fiber cable, because light is
emitted from the light source port. Visual impairment may occur if the light enters the eye.
• Close the covers of any light source ports that do not have optical fiber cables connected
to them. On models with two or more light source ports, visual impairment may occur if light
that is mistakenly emitted from the wrong port enters the eye.
Connect one end of the optical fiber or line that you need to perform loss testing on to the optical
power measurement port of the power meter, and connect the other end to the light source port of the
light source.
Executes an auto loss test
The light source generates the specified
wavelengths of measurement light in order.
The optical power meter measures the optical
power of the light that it receives.