XO FLOW – Installation instructions
65
3)
Select the module position in the drop-down menu.
4)
Adjust the minimum drive air flow using the slider so that the pressure measured using the test
manometer equals 1,5 bar.
5)
Adjust the maximum drive air flow using the slider so that the pressure measured using the test
manometer equals 3,5 bar.
6)
Adjust the anti-retraction air using the slider so that there is a low amount of airflow through
the turbine handpiece after deactivating the turbine. The pressure should not allow the turbine
to keep spinning after deactivation.
7)
Perform the adjustments mentioned above individually for each turbine and micromotor
suspension.
A drive air range 1,5 to 3,5 bars is appropriate for most turbines.
In some cases however a different interval should be used. Please see the documentation
for the air instrument(s) to be used.
Different air instruments (e.g. turbines) operate at different pressure settings. To get the best
performance in a scenario where multiple different air instruments are used, the pressure settings
must be adjusted to each type of air instrument.
The pressure settings can also be adjusted individually on a per-preset basis as an advanced option.
6.9.6
Adjusting the micromotor cooling air
To adjust the micro motor cooling air pressure setting, you will need a test flowmeter (FA-400).
1)
Attach the test flowmeter to the micromotor.
2)
Tap “micromotors”.
3)
Adjust the micromotor cooling air using the slider to 10 l/min.
6.9.7
XO ODONTOSCALER for ACTEON
®
compatible tips
If the unit is fitted with a handpiece for ACTEON
®
compatible tips (marked with ), the scaler presets
should be modified to match this handpiece and tips (marked with an “S” – for example 1US) using the
1)
Tap the user settings app.
2)
Tap “Presets/Scaler – XO ODONTOSCALER” and choose ACTEON
®
.
6.9.8
Set up the e-mail address of technical service
Tap the “Mail” app and insert the mail address that can be used to access your technical service team.
6.9.9
Adjusting cuspidor water flow
Water flow for the cup filler and cuspidor are pre-adjusted at the factory.
However, the water flow for the cuspidor can be adjusted with the regulator shown in Figure 95.
Содержание FLOW
Страница 1: ...XO FLOW INSTALLATION INSTRUCTIONS...
Страница 2: ...2...
Страница 5: ...XO FLOW Installation instructions 5 10 One month follow up 71 11 Fuses 72 12 Symbols 72...
Страница 17: ...XO FLOW Installation instructions 17 Figure 15 Lifting the unit stand Figure 16 Adjusting the unit base...
Страница 35: ...XO FLOW Installation instructions 35 Figure 43 Fitting of lower outer cylinder...
Страница 38: ...XO FLOW Installation instructions 38 Figure 48 Positioning of the cables Figure 49 Securing the cables...
Страница 44: ...XO FLOW Installation instructions 44 Figure 58 Fitting the brake Figure 59 Fitting the cover plug...