XO FLOW – Installation instructions
47
Figure 62 – Network connection
By default, the unit is configured to use DHCP to obtain an IP address on the network. If another
configuration is needed, this can be configured from the Windows desktop of the unit. The Windows
desktop can be accessed through the technical settings app.
5.12
Fit the cylindric covers
Fit the long cylindric cover using the 4 screws and fit the short cylindric cover just by pressing it into
position (only units without x-ray adaptor).
5.13
Suction hoses
Unpack the suction hoses and connect them in their designated places: the high-volume suction is
positioned furthest away from the chair and next the saliva suction hose. See Figure 63 –.
Check that suction hoses in correct length are fitted. See Instructions for use!