NB: If the ceiling height is lower than 255 cm see section 5.9.2.
3.
Turn the nut clockwise without turning the light arm post – the nut will cause the light arm post
and the bridge arm post to engage. Rotate the light arm post with the stop screw at the top of the
post, pointing towards the chairside.
4.
When the nut begins to tighten, turn the light arm post slowly until the two posts engage. Then
tighten the nut even further by hand, approx. 1 turn. Finally, fasten the nut with the supplied
special wrench (MG-416). See Figure 42.
5.8
Fitting the Navigator
The outer Ø100 mm cylinder is divided into two parts (an upper short and a lower long one) and the
light arm post has separate friction adjustments for the Navigator arm and operating light arm.
1.
Fit the light arm post as described in section 5.7.
2.
When the light post has been tightened securely, the lower Ø 100 mm outer cylinder shall be fitted
– see Figure 43.
3.
Fit the Navigator arm – see Figure 44.
4.
Route the cables from the Navigator through the groove in the plastic ring as shown in Figure 46.
5.
Fit the O-ring and the adjustable plastic ring (19 in Figure 45) on the light post over the display
arm, as shown in Figure 44.
6.
Route the cables from the Navigator through the hole in the light post. See Figure 47.
7.
Push the cables through the post until they appear in the electronics side of the unit stand.
8.
Check the movability of the Navigator arm and make sure that the cables do not obstruct the arm
movements. The cables should be positioned like a small arc to allow for movement of the arm and
secured under the non-adjustable plastic ring. The cables should furthermore be secured to the
post with a cable tie. Please see Figure 48 & Figure 49.
9.
Fit the O-ring and the non-adjustable plastic ring at the top of the light post.
10.
Fit the stop screw for light arm in the threaded hole above the non-adjustable plastic ring.
11.
Fit the wide-collar plastic ring in the groove above the stop screw, above which two bearing rings
should be placed.
12.
Fit the narrow-collar plastic ring on top of the bearing rings and secure them with an O-ring.
13.
Fit the upper Ø 150 mm outer cylinder and check that there is equal space between the lower outer
cylinder/screen arm and the upper outer cylinder/screen arm. If necessary, adjust the adjustable
plastic rings.
14.
Fit the friction brake in the rear arm of the display by turning the lower outer cylinder until the
threaded hole aligns with the hole in the outer cylinder. See Figure 51.
15.
Adjust the friction brake, following which the outer cylinder should be turned until the hole is
pointing backwards.
16.
Fit the cover plug into the hole of the outer cylinder. See Figure 59.
17.
Route the cables from the display as shown in Figure 44.
18.
Once the cables have been routed, check that the arm can rotate in its full range without pulling
the cables.
19.
Fit the operating light as described in section 5.7.
For safety reasons always remember to fit the cover plug into the hole of the outer cylinder.
See Figure 59.
Содержание FLOW
Страница 1: ...XO FLOW INSTALLATION INSTRUCTIONS...
Страница 2: ...2...
Страница 5: ...XO FLOW Installation instructions 5 10 One month follow up 71 11 Fuses 72 12 Symbols 72...
Страница 17: ...XO FLOW Installation instructions 17 Figure 15 Lifting the unit stand Figure 16 Adjusting the unit base...
Страница 35: ...XO FLOW Installation instructions 35 Figure 43 Fitting of lower outer cylinder...
Страница 38: ...XO FLOW Installation instructions 38 Figure 48 Positioning of the cables Figure 49 Securing the cables...
Страница 44: ...XO FLOW Installation instructions 44 Figure 58 Fitting the brake Figure 59 Fitting the cover plug...