6.4
Adjusting the friction brakes in the operating light arm and navigator arm
The arm for the operating light and the arm for the Navigator has three pivoting joints. The Navigator
screen have one pivot joint. Each joint has its own friction brake, which can be adjusted with Allen
keys.
Turn brakes clockwise to increase friction.
Turn brakes counterclockwise to decrease friction.
Figure 91 – Adjusting the vertical friction of the arm using a 3 mm Allen key
Figure 92 – Adjusting the friction in the light head using a 3 mm Allen key