XO FLOW – Installation instructions
32
Figure 39 – Maximum arm reach of intraoral x-ray unit
For further instructions, please refer to the instructions supplied by the x-ray unit manufacturer.
The weight of the x-ray unit head may not exceed 8 kg, and the weight of the arm system
may not exceed 17 kg.
The arm reach of the x-ray unit may not exceed 220 cm. See Figure 39.
An x-ray unit must only be installed if the unit is bolted directly to the floor. An x-ray unit
must not be installed if the XO unit is fitted with a steel installation plate as the XO unit
may not be stable!"
The x-ray unit must not be electrically connected to the XO unit!
5.7
Fitting the light arm post
1.
Fit the large nut to the bridge arm post (only until engaged by thread) by turning it on to the
bridge arm post. See Figure 40.
Figure 40 – Fitting the nut onto the bridge arm post