USER GUIDE |
EAGLE TWO+
Version: 3.3
89
e)
I-max for Saline Low & High Cut is adjusted.
Sub step- 6 (Saline Resection Coag Mode)
Equipment:
Monopolar footswitch.
Small test cables with banana plugs
True RMS current meter
10 Ω non-inductive power resistor.
Procedure:
1) Connect:
a) One test cable from Saline Resection plug to 10 Ω. (Refer the connection diagram for
Saline Resection).
b) One test cable from Saline Resection socket to true RMS current meter in series with 10Ω.
c) Monopolar footswitch to the Monopolar 1 footswitch socket on the rear panel.
2) Check and adjust the I-max for Saline Resection Coag output.
a) Select mode „2:6 BiTUR Coag.
a) Press the Monopolar footswitch pedal and check if the current reading is equivalent to 3872
mA ± 20 mA RMS.
b) If not then stop activation. If the output current is high decrease DAC Counts by pressing
down key. If low increase it by pressing up key.
c) Repeat step (b) & (c) sequentially until the meter reading is not in the stated range.
d) When reading is in stated range, press „Write Counts to program memory‟ for storing
corresponding fire count reading in the system memory.
e)
I-max for Saline Coag Soft and Force mode is adjusted.
Sub step- 7 (Sealer Mode)
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...