USER GUIDE |
EAGLE TWO+
Version: 3.3
86
f)
I-max for all the bipolar modes will be set.
3)
Disconnect the test cables from bipolar output.
Sub step- 2 (Pure Cut Mode)
Equipment:
Monopolar footswitch.
Small test cables with banana plugs
True RMS current meter
200 Ω non-inductive power resistor.
PREM plug modified.
Procedure:
1) Connect:
a) One test cable from PREM socket to 200 Ω resistor. (Refer the connection diagram for
Monopolar).
b) One test cable from Monopolar 1 socket to true RMS current meter in series with 200 Ω
resistor.
c) Monopolar footswitch to the Monopolar 1 footswitch socket on the rear panel.
2) Check and adjust the I-max for Monopolar cut output.
a) Select mode „2:2 Pure Cut‟.
a) Press the Monopolar cut footswitch pedal and check if the current is equivalent to 1224mA
± 20 mA RMS.
b) If not then stop activation. If the output current is high decrease DAC Counts by pressing
down key. If low increase it by pressing up touch key.
c)
Repeat step (b) & (c) sequentially until the meter reading is not in the stated range.