USER GUIDE |
EAGLE TWO+
Version: 3.3
77
CHAPTER 9
Testing Procedures
Warning-
Electric Shock Hazard - the generator power cord must be connected only with a
properly grounded socket. Do not use power plug adaptors.
Warning-
Fire Hazard - Do not use extension cords.
Caution
-
Do not stack equipment on top of the EAGLE TWO+or do not place the generator on top
of other electrical equipment. These configurations do not allow for adequate cooling.
Caution
-
Place the generator on any stable or flat surface, such as a table or Platform.
Carts with conductive wheels are recommended. Refer to the procedures for your
institution or to local codes for details.
Provide at least four to six inches of space from the sides and top of the generator
for cooling. Normally, the top, sides, and rear panel are warm when the generator is
used continuously for extended periods of time.
Caution
-
According to the procedures an equipotential grounding cable to the grounding plug
on the rear panel of the generator. Then, connect the cable to earth ground.
Caution
-
If required connect the generator to the hospital equalization connector with
an equipotential cable.
Caution
-
Connect the power cord to a wall socket having the correct voltage. Otherwise, product
damage may result.
Caution
-
Plug the generator power cord into a grounded socket. Grasp the plug, not the power
cord. Do not pull the cord itself.
Warning
-
Keep test leads to the minimum length usable; lead inductance and stray capacitance can
adversely affect readings. Carefully select suitable ground points to avoid ground loop
error in measurements. The accuracy of most RF instruments is approximately 1-5% of
full scale. Do not use uncompensated scope probe, causes large errors when measuring
high voltage RF waveforms.
Periodic Safety Check:
Perform the following safety check every two years to verify that the EAGLE TWO+Generator is
functioning properly. Record the test results for reference in future tests. If the generator fails to meet
any of the checks, refer to Chapter 10, troubleshooting.
Warning -
Electric Shock Hazard - When taking measurements or troubleshooting the generator,
take appropriate precautions, such as using isolated tools and
equipment, using the “one hand rule,” etc.
Warning -
Electric Shock Hazard- Do not touch any exposed wiring or conductive
surfaces while the generator is disassembled and energized.
Warning -
Electric Shock Hazard- Never wear a grounding strap when working on an energized
generator.
Caution
- The generator contains electrostatic- sensitive components. When repairing the
generator, work at a static control workstation. Wear a grounding strap when handling
electrostatic- sensitive components, except when working on an energized generator.
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...