Note down the error code and type of error and take necessary measures to eliminate as described in
the following table.
Error Code
No.
Description
Recommended Action
501
Up key pressed
Turn OFF and again turn ON the generator.
Do not press any activation during self-test.
If the alarm repeats, disconnect all
accessories. Turn OFF and again turn ON
the generator.
If the alarm repeats, check activation
request generator circuit.
Call
Whittemore Enterprises
Inc.
Pvt.
Ltd.
service cell.
502
Down key pressed
503
Enter key pressed
504
Left key pressed
505
Right key pressed
601
FootSwitch1/Handswitch1:
Cut
activation error
Turn OFF and again turn ON the generator.
Do not press any Key during self-test.
If the alarm repeats, disconnect all
accessories. Check/connect the ribbon
cable that connects the Display board to the
front panel. Turn OFF and again turn ON
the generator.
If the alarm repeats, replace front keypad.
Call
Whittemore Enterprises
Inc.
Pvt.
Ltd.
service cell.
602
FootSwitch1/Handswitch1
Co-
agulation activation error
603
FootSwitch2/Handswitch2
Cut
activation error
604
FootSwitch2/Handswitch2
Co-
agulation activation error
605
Single Paddle Bipolar/Sealer
footswitch error
606
Sealer Handswitch activation
error.
701
Temperature High signal.
Turn OFF and again turn ON the generator.
If the alarm repeats, Check working of
internal cooling fan.
Replace internal cooling fan
Call
Whittemore Enterprises
Inc.
Pvt.
Ltd.
service cell.
Sealer Warnings:-
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...