USER GUIDE |
EAGLE TWO+
Version: 3.3
99
CHAPTER 11
Care, Storage and Disposal
After each use, perform the following cleaning procedures immediately. If cleaning is delayed, debris
encrustation may become a source of infection. Encrustation may also result in electrosurgical unit
malfunction.
11.1
CARE OF THE UNIT
WARNING
After cleaning the electrosurgical unit, dry it thoroughly before using it again. If it is used when wet,
there is the risk of an electric shock.
CAUTION
Never immerse the electrosurgical unit in water, clean or disinfect by immersion, gas sterilization or
autoclaving. It may cause equipment damage.
11.2
STORAGE OF THE UNIT
Do not wipe the external surface with hard or abrasive wiping material. The surface will be scratched.
1.
Turn the electrosurgical unit OFF and disconnect the power cord from the receptacle (wall mains
outlet)
2.
If the equipment is soiled with blood or other potentially infectious materials, first wipe off all
gross debris using neutral detergent, then wipe its surface with a lint-free cloth moistened with a
surface disinfectant.
3.
To remove dust, dirt and non-patient debris, wipe the electrosurgical unit and foot switch using a
soft, lint-free cloth moistened with 70% ethyl or isopropyl alcohol.
4.
Make sure that the electrosurgical unit and foot switch are completely dry before storage.
WARNING
Do not store these devices in humid and unventilated environment as it may encourage the growth of
micro-organisms and pose an infection control risk.
CAUTION
Do not store the electrosurgical unit in a location exposed to direct sunlight, x-ray, radioactivity,
liquids or strong electromagnetic radiation (e.g. near microwave medical treatment equipment, short
wave medical treatment equipment, MRI equipment, radio or mobile phones). Damage to the
electrosurgical unit may result.
Do not apply excessive bending, straining or squeezing force to any cords during storage. It may cause
malfunction.
1. Disconnect the power cord
2. Store the equipment at room temperature in the horizontal position in a clean, dry and
stable location.
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...