SETTINGS FOR SURGERIES:
a)
Open Heart/CABG/Thoracic Surgery –
Cut Pure: 1, Coag Fulgar: 40, Bip Macro: 1
b)
Onco/Gynac /Fatty Tissues –
Cut Pure: 35-45, Coag Spray: 40-50, Bip Macro: 15 to 25
c)
Resection/ TCRE/under water Surgery –
Cut Pure: 90-120, Coag Fulgurate: 50-70, Bip Macro: 1
d)
ERCP/ Papillotomy/ Polypectomy Surgery-
EndoCut: 30-50, Coag Fulgurate: 30-50, Bip Macro: 1
e)
Plastic/ Neuro/ ENT/ Spine/Ophthal Surgery-
Cut Pure: 30, Coag Soft: 30, Bip Micro: 15-25
f)
Lapro/GI/Ortho Surgery-
Cut Pure: 30-40, Coag Fulgar: 35-45, Bip Macro:25-45
g)
Open/Hernia/General Surgery –
Cut Pure: 35-45, Coag Fulgurate: 35-45, Bip Macro: 15 to 25
E.
ALARM CONDITIONS:
Non - PREM Patient Return Electrode Alarm:
When a non - PREM patient return electrode is connected and generator detects a cord fault condition,
the PREM Alarm indicator illuminates red. When you activate in alarm condition, the indicator blinks.
The following conditions can generate a PREM alarm:
1.
The patient return electrode is not connected to the generator when the generator is activated
for monopolar surgery.
2.
The return electrode does not have adequate contact with the patient.
3.
The contact area is reduced due to movement, loss of adhesion, fluid pooling or dry gel.
4.
The return electrode cord is damaged, causing excessive resistance.
System Alarm:
When the generator senses a system alarm condition, an alarm tone sounds and the generator is
deactivated. An alarm number flashes in the cut display on the front panel.
1.
Turn off the generator.
2.
Turn on the generator & verify that the self – test is completed successfully.
If you are unable to correct the system alarm condition use standby generator to complete the surgical
procedure.
Vessel Sealer Completion Alarm
System generates completion alarm once tissue is adequately sealed. Do not remove sealing
instrument from tissue until the completion alarm is sounded.
13 cycle Alarm:
This alarm is generated when the held tissue is too thick to seal. Please hold less tissue and then
starting sealing again.
Saline Resection Alarms
System generates completion alarm on open condition, when tissue is not in proper contact with the
saline instrument.
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...