USER GUIDE |
EAGLE TWO+
Version: 3.3
41
c.
The PREM Alarm indicator illuminates red. If the self-test is successful, connect the
Accessories and set the generator controls.
C.
PROGRAMMING THE GENERATOR
Programming of User Programs
: Please follow user programs guide.
(Chapter 3 –subsection F)
D.
PATIENT RETURN ELECTRODE MONITORING (PREM)
PREM electrode is used in monopolar surgery modes. Always ensure that
the contact between Patient skin & Patient Return Electrode is adequate.
Bar below PREM icon indicates the contact adequacy.
Icon
Contact Quality
Excellent
100% Contact with patient.
Safe
Contact with patient is proper and safe.
Moderate
Contact with patient is moderate & viable. If
possible try to improve the contact, connect
PREM dual area plate properly. Take appropriate
measures to improve the contact quality.
Poor
It‟s not recommended to continue with the
surgery; Check PREM connection or replace the
notPREM electrodes.
No contact
PREM patient plate is not connected or is
damaged.
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...