Handle circuit boards by their nonconductive edges. Use an antistatic container for
transport of electrostatic – sensitive components and circuit boards.
The summary of safety checks:
Inspect the generator and accessories.
Inspect the internal components.
Test the generator.
Verify PREM Function.
Confirm outputs.
Check leakage current and ground resistance.
Recommended Test Equipment
:
Digital voltmeter
True RMS voltmeter
Oscilloscope
Leakage current tester
Leakage table – per IEC 60601-2-2,
100, 200, 300, 500 Ω, all 250 watt, 1% tolerance, noninductive
Inspecting the Generator and Accessories:
Equipments
Bipolar & Monopolar footswitch
Bipolar instrument cords (handswitching and footswitching)
Monopolar instrument cords (handswitching and footswitching)
1.
Turn OFF (0) the generator by pressing the front panel power switch.
2.
Disconnect the power cord from the wall socket.
Rear Panel
1.
Check the rear panel footswitch sockets for obstructions or damage. Check for a secure fit by
inserting the bipolar footswitch or monopolar footswitch connector into respective socket.
2.
Remove the fuse and verify correct voltage and current rating.
3.
If either connection is loose, replace the footswitch board.
Front Panel
1.
Check the bipolar instrument socket for obstructions or damage. Insert the bipolar instrument
connector (footswitching and handswitching) into the appropriate socket to verify a secure fit.
If the connection is loose, replace the front panel assembly.
2.
Check the monopolar instrument sockets for obstructions or damage. Insert the monopolar
instrument connector (footswitching and handswitching) into the appropriate socket to verify a
secure fit. If any of the connections are loose, replace the front panel assembly.
3.
Check the patient return electrode socket for a broken pin or an obstruction. If the socket is
damaged or obstructed, replace the front panel assembly.
Footswitches
1.
Remove the footswitches from the generator.
2.
Disassemble the footswitches connector. Inspect the connector for damage or corrosion.
3.
Reassemble the footswitches connector.
4.
Inspect the footswitches for damage.
5.
Reconnect the footswitches to the generator.
Power cord
1.
Remove the power cord from the unit and ensure that it is unplugged from the wall socket.
2.
Inspect the power cord for damage.
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...