Remove all gross matter (blood, mucus, tissue etc.) by wiping the PK Connector
Prepare the connector cable for steam sterilization according to the following cleaning
procedure:
Remove all the residues of blood, tissue and other objects by wiping (do not use abrasive
material) the Connector Cable with a cloth in a mild cleaning solution or detergent capable of
removing organic deposits. Do not use reprocessing solutions.
Rinse thoroughly in running water.
Remove any residues further by cleaning agents with a water damp cloth.
Dry the device thoroughly before sterilizing. Sterilize the connector cable according to the
sterilization procedure.