USER GUIDE |
EAGLE TWO+
Version: 3.3
44
Foot Switch Selection for
Eagle 2+
:
Default Configurations:-
Every time generator is turned ON or restarted, machine loads by default settings for footswitching
sockets.
For Eagle
2+ only
Monopolar 1
Footswitch
Monopolar 2 /Bipolar
Footswitch
Bipolar/Sealer
Footswitch
Can be
Used for
Cut / Coag on
Monopolar 1
Cut / Coag on
Monopolar 2
-------OR-------
Bipolar & Sealer
Bipolar
-------OR-------
Sealer
By
default
Cut & Coag on
Monopolar 1
Cut & Coag on
From Monopolar 2
Bipolar
Please Note:
For using these footswitching input sockets for functions other than default, user must manually
configure it.
When using footswitching electrodes user must connect footswitches to corresponding ports on the
rear panel.
Monopolar
As in this example if user intends to connect footswitching pencil in Monopolar -2 socket and
Handswitching electrode in Monopolar-1 socket then user must connect double paddle Footswitching
electrode in Monopolar-2 footswitching socket only on the rear panel, as shown in figure below.
Monopolar 2
OR
Sealer &
Bipolar
Monopolar 1
Sealer
OR
Bipolar
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...