Vessel Sealing Instrument Guidelines
BEFORE SURGERY:
Check all parts of Outer tubing, Blade assembly, Jaws and Insert assembly and Handle for
mechanical deformities and insulation breakages. Open the Outer tubing & Blade assembly to
check the proper cleaning of the instrument.
Clean the Jaws & Hinge areas by using H
2
O
2
followed by any permissible disinfectant as per
the operation theatres norms.
Make sure that there is no tissue left inside the hinge areas & blades assembly tubing. This can
damage the instrument during next use.
In case of using non sterilized blades, assemble the instrument with blade earlier to packing the
instrument for sterilization.
Cover all the parts and assembly separately in the cloth as open assembly may damage the
insulation of other parts during sterilization.
Pack all covered assembly in one bunch and send it for sterilization.
DURING SURGERY:
Blade is an incisive element. Always handle the blade cautiously.
Do not use Normal Saline (NS) to clean the instrument at any stage. Usage of Saline could
result in failure of the instrument.
Do not activate the blade without completion of sealing cycle as this may cut the live arteries
resulting to blood loss during surgery.
Prepare the cleaning solution with 50% H
2
O
2
& 50% permissible disinfectant (like Betadine)
for cleaning the fat and tissue deposition on jaws during surgery.
DO NOT USE NORMAL SALINE (NS) FOR CLEANNING OF THE INSTRUMENT DURING THE
SURGERY
Open the sterilized bunch of the instrument in sterilized field in OT (Operation Theatre); open
the cloth cover of all the assembly.
Inspect all assembly for physical damage and insulation failure during sterilization.
Assemble the instrument as per the assembly instruction and check the functioning before
using it in surgery.
Do the periodic cleaning of jaws by the preparing cleaning solution 50% H
2
O
2
& 50%
permissible disinfectant.
In case of incomplete sealing or alarm condition, do not activate the blade.
Sealing instrument may not be suitable for sealing the arteries below 1mm due to functional
behavior. Use other instrument in such case.
Do not use the device in blood reach environment & do not give wash of Saline till procedure
completes; as vessel sealing do not function in conductive medium.
In case of open re-hold (OP re Hold) alarm, check instrument assembly & the cable
connection, also hold the tissue properly in the device to restart the sealing cycle. If this
problem persists, please call us for the technical help.
In case of close re-hold (CL re Hold) alarm, check the instrument assembly, any metal object/
clips in the jaw during sealing
or
short circuit in jaws
or
usage of device in presence of blood
reach environment or normal saline (NS). Rectify the condition & retry sealing cycle till
completion alarm sounds. If this problem persists, please call us for the technical help.
Do not hold more than 7 mm tissue chunk for sealing as this may lead to increase in sealing
cycle & lateral thermal spreading.
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...