USER GUIDE |
EAGLE TWO+
Version: 3.3
5
Bipolar Cut & Coag with independent Display setting with independent foot-switches.
Sealing Program auto-select is available as per the selected accessory.
SECTION 3: Saline Plasma Bipolar Resection
Electrode Disconnection / Non Saline activation Alarm.
Regular loops can be used for saline work, hence reduces working cost.
Regular Working Element can be used for bipolar work.
B.
MODES OF ELECTROSURGERY
1.
Monopolar Electrosurgery:
In Monopolar electrosurgery, single pole, active electrode is in the surgical site. An electric current
from generator is delivered to the surgical site through active electrode & returned back to generator
via patient return electrode. Monopolar electrosurgery is used for most of the general surgical
procedures. Patient return electrode is used in these applications as return path for RF current.
Monopolar
Cut
Modes
: There are SEVEN Cut modes-
Low Cut
-
It is used for a cut with no sparking, useful for precise and clean cutting for thin
tissue.
Pure
-
This is the default Monopolar cut mode it may be used for precise cut with no
hemostasis.
Blend 1
-
It is used where slower cutting & moderate hemostasis is desired.
Blend 2-
It is used where slower cutting & extra hemostasis is desired.
Endo-1 Cut
-
It is used for a cut with pauses & moderate pulse interval control.
Endo-2 Cut
-
It is used for a cut with pauses & large pulse interval control.
Hydro Cut –
It is used for efficient cutting under bodily fluids/water.
Monopolar
Coag
Modes
: There are SEVEN Coag modes-
Soft
–
Soft coagulation with deep tissue necrosis.
Swift -
Low voltage non-contact coagulation
Desiccate
–
Coagulation for soft and delicate tissues.
Fulgurate
–
This mode is suitable for long remote/non-contact coagulation with moderate
sparking in variety of application. This mode can be selected separately as and when required.
Spray
–
This mode is suitable for long remote/non-contact coagulation with moderate
coverage.
Hi-Spray
–
Non-touch long distance coagulation with larger coverage area.
Hydro-Coag
–
It is used for efficient coagulation under bodily fluids/water.
2.
Bipolar Electrosurgery:
In Bipolar electrosurgery, both the active & return electrodes are in surgical site. Patient return
electrode is not required as bipolar instrument contains an active electrode & return electrode.
An electric current flows from active electrode to the return electrode through tissue grasped by the
instrument. This technique is mostly used in delicate surgeries, cosmetic surgeries & neurosurgeries.
Bipolar Modes
: Bipolar Electrosurgery contains FIVE types of modes-
Micro
–
The voltage is kept low to avoid sparking. It is selected for fine bipolar tissue
desiccation.
Standard
–
Standard bipolar coagulation for all type of tissue with moderate hemostatis.
Force
–
Bipolar coagulation for thicker & high impedance tissue.
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...