USER GUIDE |
EAGLE TWO+
Version: 3.3
106
Sealer Open warnings:-
Causes:-
Sealer clamp is not applied / not properly applied on a vessel when sealer is activated.
Connector is not properly connected in the sealer port.
Sealer clamp wire breakage.
Measures:-
Clean sealer clamp‟s head clip openings properly.
Hold sealer clamp properly & apply adequate pressure on tissue using sealer clamp.
Check continuity of individual clamp wire using multimeter. If it‟s open change sealer cables.
In case, if problem still exists then contact Whittemore Enterprises Inc.
Sealer Close warnings:-
Causes:-
Sealer clamp is activated on conducting material/ in conducting fluid.
Connector is not properly connected in the sealer port.
Sealer clamp cable is shorted internally.
Sealer clamp head opening is bent or damages due to excessive external pressure from clamp
head opening is not cleaned properly.
Measures:-
Clean sealer clamp‟s head clip openings properly.
Keep sealer clamp away from vicinity of conducting material or fluids.
Hold sealer clamp properly & apply adequate pressure on tissue using sealer clamp.
Check continuity between two wire using multimeter. If shorted replace the sealer cable.
In case, if problem still exists then contact
Whittemore Enterprises
Inc.
Содержание EAGLE TWO PLUS
Страница 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Страница 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Страница 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Страница 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Страница 49: ...USER GUIDE EAGLE TWO Version 3 3 48 2 Electrosurgery Mode in Eagle 2 Accessories connection in ESU Mode ...
Страница 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...