b)
Press the Monopolar coag footswitch pedal and check the current equivalent to 89 mA ± 5
mA RMS.
c)
Stop activation. If the output current is high decrease DAC Counts by pressing down key.
If low increase it by pressing up key.
d)
Repeat step (b) & (c) sequentially until the meter reading is not in the stated range.
e)
When reading is in stated range, press „Write Counts to program memory‟ for storing
corresponding fire count reading in the system memory.
3) Disconnect the test cables from Monopolar output.
4) Use same equipments & repeat the same procedure for calibrating ‘Vpk’ of different
modes (Desiccate/ Swift/ Fulgar/ Spray/ Hi-Spray) by choosing corresponding calibration
modes.