C H A P T E R 6
–
S A F E T Y S Y S T E M S
AT5600 User Manual 98-119 issue 14
Page 87
6.6.
Constructing Safe Fixtures
When constructing a test fixture, the external surfaces of the transformer and its
connections, must
NOT
protrude beyond any of the edges of the test fixture at
any time
This is to ensure that a safe working distance of 100mm minimum, is maintained
from the transformer under test to the Safety Light Curtain Beam, which is
protecting the operator(s) from the dangers of electric shock.
This ensures that the safety system has had time to de-energize the component
under test before the user's hand has reached the component or it's connections.
6.7.
DC1000 DC Bias Current Source
When operating the DC1000A(s) with the AT5600, For Safety Requirements refer
to DC1000A User Manual (98-102).
6.8.
AC Interface Fixture
When operating the AC Interface Fixture with the AT5600,
For Safety Requirements refer to AC Interface Fixture User Manual (98-072).
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...