After you have clicked on the ‘OK’ button, the tester will be commanded to make
its measurements, and you will see the following message:
When the measurements have finished, the display will return to the dialogue box
for the LS test the measured nominal inductance:
From this point, it is very easy for you to complete the dialogue box by simply
entering the values you require for the Negative and Positive percentage limits.
If you have chosen a fixed voltage, current or frequency conditions that may not
be possible for the tester to make a measurement. For example, if the AC
impedance of the winding under test is only 0.1
at the specified frequency and
you have specified a 5V test signal, then the AT would have to supply 5V / 0.1
=
50 Amps!
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...