C H A P T E R 1 4 - A T E D I T O R
AT5600 User Manual 98-119 issue 14
Page 370
To program the test:
1. Enter the value of a voltage (or current) for the signal and the required
frequency of the test.
Also, enter the bias current you would like to apply.
2. Select the integration time required.
‘Medium’ is the default setting. ‘Long’ will give more stable readings (for
tighter limits) at the expense of time, and ‘Short’ will test at the maximum
speed, but may give slightly noisier readings. See 7.3 Explanation of
Integration.
3. Enter the terminal names between which the impedance is to be
measured.
4. Set the impedance limits.
+/-
Enter a nominal value and then the limits as negative and
positive values.
><
Enter minimum and maximum values.
>
Enter just a minimum value.
<
Enter just a maximum value.
The Measure button may be used here as outlined in section 14.5.9.
5. Select OK. The test specification will then be displayed in the program
window.
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...