C H A P T E R 7
–
T E S T S A N D T E S T C O N D I T I O N S
AT5600 User Manual 98-119 issue 14
Page 114
Best Practice for Centre Taps / AutoTransformers (also applies
to LVOC)
When testing TR on windings where the energized/primary are on the same coil
as the secondary, it is important to minimize the effect of common mode signals
to get the best measurement.
This can be achieved by programming the energized nodes, so that the
secondary winding is always at the LO end of the primary. As the AT EDITOR
allows you to choose any node it is best to consider this when programming the
test.
Please consider the following two examples which illustrate this point.
A, Best Practice set up
All the LO terminals
are defined as “A”
The secondary is at the LO end of A-C ; SEC LO is at 0V
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...