During a Hi-Pot test, the Wattage is continuously monitored and adjusted to
provide the correct output. If the programmed output voltage cannot be
maintained, the tester will automatically display ‘FAIL’, illuminate the red indicator,
return a ‘FAIL’ to
the server (if used) and sound a warning buzzer (if enabled).
This method complies with the production requirements of EN61558-1 (1998),
UL1411, 5
th
Edition and others.
Please also see
14.6.25 for specific advice on programming the test, and
10.2.13 for test accuracy specification