9.2.
Correcting Errors
9.2.1.
Safety Interlock error
The safety interlock system must employ all three control lines and if they are
not all being switched correctly then the safety interlock icon will turn RED,
.
See section 10.3.5 for details of the safety interlock interface.
9.2.2.
Temperature error
If the AT5600 gets too hot, then an error message indicating an Over
Temperature shall be displayed in the error message window.
Overheating can be caused by many factors;
1. Clogged air filters
2. Blocked air vents
3. Ambient temperature too high
4. An internal fault
If a temperature error is shown on screen or in any Measurement Status Codes,
then make sure adequate space is allowed around the air vents (see 2.6) and
clean the air filters as shown in section 11.1.
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...