C H A P T E R 1 0
–
S P E C I F I C A T I O N S
AT5600 User Manual 98-119 issue 14
Page 211
10.2.3.
QL and D Tests
The accuracy is based on the chosen test conditions, and is given by:
𝐴
𝑇
= 𝐴
𝑅
+ 𝐴
𝐼
+ 𝐴
𝑄𝐷
where:
A
T
is the total accuracy (%)
A
R
is the basic relative accuracy (%)
A
I
is the correction for current level (%)
A
QD
is the correction for Q or D factor (%)
where:
𝐴
𝑅
= 0.50 %
𝐴
𝐼
= 0.08 % +
0.0001 %
𝐼
𝑀
𝐴
𝑄𝐷
= 0.2 % ∗ [𝑄
𝑀
+ (
1
𝑄
𝑀
)] (𝐹𝑜𝑟 𝑄 𝑓𝑎𝑐𝑡𝑜𝑟)
𝐴
𝑄𝐷
= 0.2 % ∗ [𝐷
𝑀
+ (
1
𝐷
𝑀
)] (𝐹𝑜𝑟 𝐷 𝑓𝑎𝑐𝑡𝑜𝑟)
where:
I
M
is the measured current (A)
Q
M
is the measured Q factor
D
M
is the measured D factor
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...