C H A P T E R 7
–
T E S T S A N D T E S T C O N D I T I O N S
AT5600 User Manual 98-119 issue 14
Page 152
7.36.
MAGX - Magnetizing Current (External Source)
Where Used
The AT5600 offers two basic alternative ways to confirm that the transformer has
been assembled properly, with the appropriate number of primary and secondary
turns, the right grade of magnetic material for the core, and the correct air gap if
required.
Magnetizing current and open circuit voltage are the preferred tests for line
frequency transformers, designed to operate over the full extent of the B-H curve,
including the non-linear regions.
(For other transformers, such as pulse transformers and those used in switched
mode power supplies, inductance and turns ratio are the preferred tests.)
This version of the MAGI test uses an external source to provide the test signal
which must be coupled to the tester via a Voltech AC Source Interface (contact
your supplier for details). See the AC Source Interface user manual or the Editor
help system for details of how to configure external AC sources for use with an
AT5600.
Measurement Conditions
When measuring magnetizing current, you should normally program the test to
apply the highest working voltage at the lowest working frequency to the primary
winding.
In the case of a transformer with a split primary, the test can be conducted equally
well by energizing just one of the primary windings, as opposed to the two in
series. The expected current will be greater for the single winding, rising in
proportion to the turns ratio:
I
A
= I
AB
x (N
AB
/ N
A
)
where
I
A
= The current to be specified when testing with winding A
I
AB
= The current for windings A and B in series
N
A
= The number of turns on winding A
N
AB
= The number of turns on A and B in series
(As an alternative, the formula above can be written using the voltage ratio
between the two windings, rather than the turns ratio.)
In principle, you may measure the magnetizing current using any winding, or any
series combination of windings, with the current limit adjusted according to the
formula above, because the Ampere-turns required to magnetize a transformer to
a given flux level is independent of which winding is used.
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...