List of input/output signals
Signal identification
Contents
Pin
No.
Signal form
-COM(ANALOG GND)
Open
collector
output
common
1 GND
SET POINT L
Leak rate low setpoint
2
Open
collector
(+24 V input)
SET POINT H
Leak rate high setpoint
3
Open
collector
(+24 V input)
START UP OK
Outputted when instrument
startup is completed
4 Open collector
(+24 V input)
SYSTEM ERROR
Outputted when trouble has
occurred in equipment
5 Open collector
(+24 V input)
Output
-0
Outputted when leak rate is
minus
6 Open collector
(+24 V input)
+COM(VOLT INPUT)
Common for input
9
DC +24 V input
AUTO ZERO
Auto zero (zero adjustment)
ON
10 Normally open
(GND)
CALIBRATION Automatic
sensitivity
calibration ON
11 Normally open
(GND)
TEST START/STOP
Start/end of test (vent)
12
Normally
open
(GND)
PRINT OUT
Printing (can print out at will)
13
Normally
open
(GND)
Input
SYSTEM START/STOP
Start/stop of system
14
Normally
open
(GND)
(2) Description of signals
<Output signal>
-COM(ANALOG GND)
→
Common for open collector output
SET POINT L
→
Outputted when the leak rate measurement value
is higher than the setpoint L value. Outputted in
test state. This output is reset at the end of test.
SET POINT H
→
Outputted when the leak rate measurement value
is higher than the setpoint L value. Outputted in
test state. This output is reset at the end of test.
START UP OK
→
Outputted from the time when startup is normally
completed and standby status is set up. This
signal is outputted until stop action is started, but
if automatic sensitivity calibration is executed, the
signal will not be outputted for the time.
SYSTEM ERROR
→
Outputted if any trouble occurs. It is reset the
moment the error mode is reset by pressing
.
(when the DIP switch signal output setting is
normal)
-0
→
Outputted when the leak rate measurement value
is minus (value smaller than zero). Outputted in
the test status.