- 2 -
1-2 Operating
Modes
HELIOT runs according to the program installed in the basic unit. Its operating modes can
roughly divided into the initial state, startup mode, test mode, stop mode and error mode.
Representative series of actions are given below.
Initial state Main power ON
The main power is turned ON.
↓
Press START/STOP.
Pumping system starting
The molecular compound pump is in static rotation.
↓
Zero adjustment
Measures the zero point (background).
↓
He peak adjustment
Searches for the most efficient voltage for helium
detection.
↓
Fine leak flow sensitivity calibration
Performs fine leak flow sensitivity calibration.
↓
Startup mode Middle leak flow sensitivity
calibration
Performs middle leak flow sensitivity calibration.
↓
Gross leak flow sensitivity calibration Performs gross leak flow sensitivity calibration.
↓
Zero adjustment
Performs zero point measurement.
↓
Test mode standby
Ready for execution of test.
↓
Press TEST
Test port roughing
Pumps the test port from atm. pressure.
↓
Test mode
Gross leak flow
Executes test of gross leak flow.
↓
Middle leak flow
Executes test of middle leak flow.
↓
Fine leak flow
Executes test of fine leak flow.
↓
Press VENT
Test mode standby
Ready for execution of test.
↓
Press START/STOP.
Stop mode
Pumping system OFF
Terminates the action of pumping system.
↓
Electrical system OFF
Turns OFF the electrical system.
↓
Main power OFF
Message
display
Displays error message
↓
Error mode On standby
Waits by displaying message.
↓
Press
Returns to test mode standby
In case of "Notice" display, returns to standby.
↓
Stop mode execution
In case of "Caution" display, executes the stop mode.