CAUTION
Quick start
Quick start starts the instrument by utilizing the sensitivity data used the
last time, as it is, independently of the current state of HELIOT.
Therefore, the measurement error can be great in some cases.
To ensure test data reliability, it is recommended to start up the instrument
by normal start.
It is to be understood that the reliability of leak test when the instrument is
started by quick start (direct reading of leak rate, etc.) is not warranted.
!
CAUTION
Test port
Check that a blank flange or a test object is securely connected to the test
port.
If test is made without connecting a blank flange or a test object securely,
not only measurement will be adversely affected, but also the life of parts
will be extremely shortened, causing damage to the equipment.
!
CAUTION
Shutdown
To shut down the instrument, set up the stop mode in the test mode standby
status for automatic shutdown.
!
CAUTION
During test
When test is going on, do not remove the blank flange or test object from
the test port or vent the test port to atmospheric pressure by other method
than operation of the controller. The equipment can be damaged.
!
CAUTION
Venting the test port
Before venting, do not remove the blank flange or test object from the test
port nor vent the test port to atmospheric pressure by other method than
operation of the controller. The equipment can be damaged.
!
CAUTION
Main power OFF
When stop operation is being performed, the message "Please wait" is
displayed on the LCD. At this time, do not turn OFF the main power.
Then the display changes to "Push START". Turn OFF the main power
after verifying this message is displayed.
!
CAUTION
After shutdown
Do not give vibration or impact to the instrument for about five minutes
after completion of shutdown operation.
The equipment can be damaged.
!
CAUTION
Main power OFF
Be sure to turn OFF the main power after completion of shutdown
operation.
!
CAUTION
Error
If error occurs, reset the error mode after verifying the symptom by the
message.
Conduct inspection and repair referring to "Troubleshooting".
!
CAUTION
Check before startup
Check again that all preliminary operations above, including installation,
DIP switch setting, SEN. switch setting and controller setting, etc., have
been properly completed.
If there is any fault or problem, not only the instrument may malfunction,
but also it may be damaged.