- 48 -
4-8-2 List of error messages
Error No.
Error message
Trouble symptom
**** Caution 1 ****
Fore pump trip
The fore pump (GHD-030) has stopped and cannot be started.
**** Caution 2 ****
Main pump trip
The main pump (TMH063-003-002) has stopped and cannot be started.
**** Caution 3 ****
Fil. burn out
Both filaments 1 and 2 have burnt out.
Emission current cannot be obtained due to deterioration or
contamination of the filament.
**** Caution 4 ****
Pig. failure
The Pirani gauge sensor head has burnt out.
**** Caution 5 ****
Ext. power trip
The external power supply has tripped.
**** Notice 1 ****
Error
Action cannot be executed for the setting.
**** Notice 2 ****
F. flow invalid
Fine leak flow cannot be executed.
(Pressure does not lower, pressure fault has occurred.)
**** Notice 3 ****
M. flow invalid
Middle leak flow cannot be executed.
(Pressure does not lower, pressure fault has occurred.)
**** Notice 4 ****
G. flow invalid
Gross leak flow cannot be executed.
(Pressure does not lower, pressure fault has occurred.)
**** Notice 5 ****
Evacuation fault
Roughing is not completed within three minutes.
**** Notice 6 ****
Sen. fault
Sensitivity fault has occurred.
**** Notice 7 ****
Fil. switched
Filament 1 or 2 has burnt out and was changed to the other.
**** Notice 8 ****
Zero fault
Zero adjustment cannot be made at startup.
RS-232C Error
Communication error of program data.
* Refer to "6. INSPECTION & MAINTENANCE" and "7. TROUBLESHOOTING" for
inspection, maintenance and replacement of parts regarding error messages.