- 30 -
Sets signal output ON/OFF to the printer. When set at ON, the instrument starts up
and the data and leak rate at the end of test (when vented) are automatically printed out
on the printer. When set at OFF, they are not printed out.
The test result is not affected by whether the printer is ON or OFF.
Printer
ON
1) Set up the printer print selection mode.
The current setting is displayed.
↓
Press
Printer
OFF
↓
2) Press
and change the setting.
ON
OFF
↓
Press
3) Press
and enter the setting.
Please wait
4) The message "Please wait" is displayed.
↓
Printer
OFF
5) The value set previously is displayed several seconds later.
This completes the input of setting.
(9)
Entering the setting of the filament change mode
The ion source used for HELIOT has two filaments. If one filament is deteriorated
and sensitivity has lowered or burns out, it is automatically changed with the other
filament. This mode permits changeover at will. Filaments are numbered 1 and 2
and either one can be used.
However, if the filament is changed over in the test standby mode after startup, test
cannot be executed unless sensitivity calibration is made by the sensitivity calibration
mode.
Change filament
filament 1
1) Set up the filament change mode.
The current setting is displayed.
↓
Press
Change filament
filament 2
↓
2) Press
and change the setting.
No.
1
No.
2
↓
Press
3) Press
and enter the setting.
Please wait
4) The message "Please wait" is displayed.
↓
Change filament
filament 2
5) The value set previously is displayed several seconds later.
This completes the input of setting.