- 46 -
Auto flow fine
1.8E-10 Pa·m
3
/s
1) Test is under way by auto flow.
↓
**** Caution 3 ****
Fil. Burn out
↓
2) Filament burnout has occurred.
The error message is displayed, the test valve is closed and the
instrument is on standby.
↓
Press
3) Press the return switch
to reset the message and standby
state.
* ULVAC HELIOT *
Please wait
↓
Several seconds later
4) The stop mode is set up.
The electrical system and pumping system are automatically turned
OFF, allowing the main power to be turned OFF.
* ULVAC HELIOT *
Push START
5) The message changes in several seconds and the message "Push
START" is displayed.
6) Shutdown action has ended. Turn OFF the main power.
(2) “Notice”
This error is displayed if any trouble occurs in a component part or malfunction occurs.
The stop mode is set up simultaneously with the resetting of the message and the
instrument is shut down or control returns to test mode standby depending on the
contents of the error. If the error occurs during test, the test will be inactive.
Specific examples are given below.
Auto flow fine
1.8E-10 Pa·m
3
/s
1) Test is under way by auto flow.
↓
----- Notice 4 -----
G. flow invalid
↓
2) Gross flow is disabled.
The error message is displayed, the test valve is closed and the
instrument is on standby.
↓
Press
↓
3) Press the return switch
and reset the message and standby
state.
Test mode L
Standby
4) Test mode standby is set up.
Test can be made again.
* If filament change occurs during test or standby, control will return to the standby mode, but test
cannot be executed unless sensitivity calibration is made after that.
Perform automatic sensitivity calibration by the sensitivity calibration execute mode by setting
the controller.
If filament change occurs during startup by normal start, sensitivity calibration will automatically
be made again after the filament is changed. In this case, the message "Fil. switched" is not
displayed.