- 44 -
4-7 Stop
Mode
This mode cools down and shuts down the instrument.
When this mode is executed, the electrical system and pumping system are automatically
stopped. Stop can be made from any mode or any state, but perform it from the test mode
standby state.
!
CAUTION
Stop
To shut down this instrument, set up the stop mode in the test mode
standby state and perform automatic stop.
!
CAUTION
Main power OFF
When stop action is being executed, the message "Please wait" is
displayed. At this time, do not turn OFF the main power.
Then the display changes to "Push START". After verifying that this
message is displayed, turn OFF the power.
4-7-1 How to shut down
Test mode L
Standby
1) Set up the test mode standby state.
↓
Press STOP
2) Press the START/STOP switch once.
* ULVAC HELIOT *
Please wait
↓
Several seconds later
3) The stop mode is set up.
The electrical system and pumping system are automatically
turned OFF and the main power can be turned OFF.
* ULVAC HELIOT *
Push START
4) The message changes in several seconds, displaying the
message "Push START".
5) Upon completion of shutdown action, turn OFF the main
power.
4-7-2 After
shutdown
The compound molecular pump is rotating at a high speed even after cool-down is completed
and the main power is turned OFF. Do not move the instrument nor give vibration or impact
for about five minutes until automatic venting of the compound molecular pump is completed
and the pump comes to a complete stop.
After shutdown, be sure to turn OFF the main power to prevent erroneous operation.
!
CAUTION
After shutdown
Do not give vibration or impact for five minutes after completion of
shutdown action.
!
CAUTION
Main power OFF
Upon completion of shutdown action, be sure to turn OFF the main power.