- 82 -
8-4 PRINTER
Output
This output prints out the startup sensitivity (minimum sensitivity) in each flow or
measurement value (leak rate) on the printer.
The recommended printer is as follows, but a printer of the same specification can be
connected.
!
CAUTION
Printer
If a printer of differing specifications is connected, the printer may not
print properly or may malfunction.
(1) Recommended
printer
Printer made by SEIKO
• Thermal printer
DPU-411 type II
• Input signal
8 bit parallel, compatible with Centronics
• No. of digits printed
40 digits
•
Size
of
character
2.47
×
1.88 mm
• Thermal paper
112 mm wide, roll 28 m long
(2)
Setting of printer print
1)
Printer print setting ON
Selecting printer print by setting the controller prints out the minimum sensitivity
in each flow at the end of startup of the instrument.
Also the test measurement value at that point (leak rate) is automatically printed
when venting at the end of test.
The test measurement value can be printed by pressing the return switch
during test.
2) Setting
printer
Setting printer print at OFF by setting of the controller does not automatically
prints startup sensitivity data or data at the end of test.
(3) Printing
1) Sensitivity
Indicates the current sensitivity mode.
Low
→→
Sensitivity L
Hi
→→
Sensitivity H
2)
Ion acc. V
Indicates the internal adjustment voltage. Not related to test data.
3) TV1
Indicates the minimum sensitivity of fine leak flow.
4) TV2
Indicates the minimum sensitivity of middle leak flow.
5) TV3
Indicates the minimum sensitivity of gross leak flow.
6) No
00001
Indicates the test result (test measurement value). Serial number from 00001
through 99999. The number is initialized by executing startup or sensitivity
calibration.
* Fore the sensitivity mode setting, refer to "4-3-2 Setting the SEN. switch".