15
YK16-0010-DI-002-02
C) Solder power supply cable to plug of power connection connector as shown in the table below.
Make sure wire size is applicable wire size.
Pump model-
CR60B-T
CR60B-S
X
R phase
L phase
Y
S phase
N phase
Z
T phase
G
PE/GND
PE/GND
Table. 5 Specifications list for connector
Pump model-
CR60B-T
CR60B-S
Receptacle maker
Receptacle part No
NET244-RM
NET243-RM
Compatible plug
NET244-PF
NET243-PF
AWG#14
AWG#12
(2.0 mm
2
)
(3.5 mm
2
)
Rated current
15A
20A
Compatible wire size
Nanaboshi Electric Mfg.Co.,Ltd.
D) Mount a guard to power connector after connection.
Fig. 10 Mounting method for guard
(1) Use supplied connectors only. If other connectors are used, first ground
contact may fail, and it causes possibility of electrical shocks.
(2) Use appropriate insulated wire, which can tolerate above rated current
value, for power cord.
(3) Do not perform insulation resistance tester.