ix
YK16-0010-DI-002-02
0.5.2 Transfer
0.5.3 Ambient Condition for Storage, Install and Operation
As precise clearances are provided with this machine, be sure to fulfill the following for its storage,
install and operation;
(1) Ambient temperature and humidity for storage
: -10°C to 60°C, less than 90%RH
(2) Ambient temperature and humidity for operation
: 15°C to 40°C, less than 80%RH
(3) Height (for both storage and operation)
: Lower than el. 1,000m.
(4) External vibration (for both storage and operation)
: Vibration acceleration less than114dB
(0.5G)
(5) Miscellaneous (for both storage and operation)
:
a. There shall be no corrosive and explosive gas.
b. There shall be no freezing or condensing.
c. There shall be no dust.
d. The pump shall be used in ventilated indoor.
e. Another pump shall not be put on the Pump.
The Pump shall not be laid down sideways.
f. There shall be no direct sun beam.
g. Heat source shall be put away from the Pump.
h. After moving the pump to place of installation, set it in a horizontal position,
i. Ensuring the ventilation space of at least 0.3m from the opening the ventilation.
Do not give the Pump a shock or lay it down. It might impair the Pump
operation.
Set it in a horizontal position keeping inlet flange upward.
CAUTION
This unit is air-cooled. Do not install directly on the ground
b
y removing the
casters. Ventilation openings under the pump are blocked, you have a risk of
getting burned or fire caused by over heat.
WARNING
(1) Although this pump is provided with casters, do not move it a long
distance using these casters.
(2) Weight of this product is as follows.
CR 60
: 48 kg
You have a risk of giving damage to your back as the load larger than
safety standard shall be required to transfer the product. Be sure to use
the loading machinery (such as mobile crane) to lift up the Pump or load it
on the pallet and fix it with Jack and run the Pallet truck for its transfer.
WARNING
If have been kept at a temperature lower than the ambient temperature is
operable,
p
lease drive temperature of the pump body from the Ambient
temperature and humidity for operation.