PhenoMaster | Hardware Operating Instructions
Tubing connections…
…must be safeguarded with tubing clamps or other safety devices, etc. to prevent them from being
accidentally loosened.
Operation
11.7.3.
•
Ensure that the operating air temperature is within the range -5...+40°C (23…104°F)
•
Ensure that humidity is <90%.
11.8. Protection Air Filter
System configuration
This section describes installation of protection air filters of CaloSys modules working with push
principle (see Section 11.1.1). For instructions of protection air filters of CaloSys modules working
with pull principle, see Section 11.1.2. When installing, ensure that filters are inserted in the correct
flow direction (i.e. reversed as compared to push principle). There is a flow direction arrow on the filter
label. Observe safety instructions regarding filter replacement intervals (compare Section 11.9.1
"CAUTION - Damage to Control Units and Air pumps! (…))"! Note that additional disk filters are
included downstream of Air OUT of animal cages (see Section 10.5).
To protect the mass flow controllers in Calorimetry Control Units against dust and particles, the
addition of protection filters between the air pump and Calorimetry Control Units is mandatory.
Connections Protection Air Filter
11.8.1.
Observe the instructions about the flow direction of the filter units in the following description.
Insert the filters into your CaloSys module by following these steps:
•
Power OFF the complete system and disconnect the air pump from the mains supply.
Содержание Calorimetry PhenoMaster
Страница 6: ...6 PhenoMaster Hardware Operating Instructions info TSE Systems com www TSE Systems com ...
Страница 13: ...CaloSys Modules 13 Confirm default installation path by pressing Next info TSE Systems com www TSE Systems com ...
Страница 15: ...CaloSys Modules 15 Press Finish info TSE Systems com www TSE Systems com ...
Страница 69: ...CaloSys Modules 69 info TSE Systems com www TSE Systems com ...