52
PhenoMaster | Hardware Operating Instructions
Fig. 25 Gas pressure display
1
Calibration gas1 pressure (Gas Inlet1)
2
Calibration gas2 pressure (Gas Inlet2)
3
Calibration gas3 pressure (Gas Inlet3)
Operating of gas pressure display buttons…
…is reserved for TSE Service only!
Fig. 26 Calibration Master Control Unit - Rear panel
1
USB connection to PC.
2
Tube connection to sample pump of the Air Drying Unit via separating bottle.
3
Tube connection to Sample IN of Calorimetry Control Unit.
4
Tube connection to calibration gas cylinders via pressure regulator.
"X0ext." connects to "X0ext." of the Calorimetry Master Control Unit. "Sample IN" connects to the
sample switch pump via Air Drying Unit. "Sample OUT" connects to the "Sample IN" of the
Calorimetry Control Unit.
For automatic calibration, the calibration gas cylinders should be connected via pressure regulator to
the "Gas Inlets 1...3" of the rear panel (see Fig. 26, [4]).
Observe instructions given below what MAX. calibration gas pressures can be applied when
calibration gas cylinders are connected to the "Gas Inlet" of the Calibration Master Unit via pressure
regulator.
•
For CaloSys modules that are operated with standard sensors of the Calorimetry Control Unit,
connect calibration gas cylinder via a pressure reducing valve set to <2bar (<29psi) above
atmospheric pressure ("Gas Inlets1...3").
•
WARNING!
Too high operating pressure could endanger persons in the vicinity (e.g. by
tubing or other materials bursting, etc.). Never connect a gas cylinder directly to an inlet
2
1
3
1
2
3
4
Содержание Calorimetry PhenoMaster
Страница 6: ...6 PhenoMaster Hardware Operating Instructions info TSE Systems com www TSE Systems com ...
Страница 13: ...CaloSys Modules 13 Confirm default installation path by pressing Next info TSE Systems com www TSE Systems com ...
Страница 15: ...CaloSys Modules 15 Press Finish info TSE Systems com www TSE Systems com ...
Страница 69: ...CaloSys Modules 69 info TSE Systems com www TSE Systems com ...