68
PhenoMaster | Hardware Operating Instructions
Appendix
A. Example configurations
See Hardware Operating Instructions - Core Modules, Appendix, "Example configurations".
B. Example electrical requirements*
8-place CaloSys-Treadmill system
Device
Power consumption [W]
Units Total power consumption [W] Voltage
Industrial PC
560
1
560
50Hz/230V
PC monitor
30
1
30
Printer
75
1
75
Air pump
150
1
150
CaloSys Control Unit
200
1
200
CaloSys Control Unit for Valves 100
1
100
Air Drying Unit
100
4
400
CaloSys Calibration Unit
100
2
200
Treadmill Control Unit
250
1
250
Treadmill
75
8
600
Total: 2565
*Under normal operation conditions. Power On peak is higher and can be calculated from max. current (fuse) and
voltage.