PhenoMaster | Hardware Operating Instructions
11.9.1.2. 4x Air Drying Unit
Fig. 37 4x Air Drying Unit
Allows for simultaneous extraction of moisture from the sample air of <=4 cages per Air Drying Unit.
Every "Sample IN / Sample OUT" pair has its own collection vessel. Every vessel has an outlet close
to the bottom that is connected to tubing.
Removal of condensate
•
The condensed water of the collection vessels must be drained off when they are half-full of
condensate at the LATEST, in order to keep the amount of moisture entering the tubing
connections as low as possible. To drain off, a clamp attached to the tubing must be opened.
Do NOT remove collection vessels!
•
The condensed water of the separating bottle must be discarded when it is half-full of
condensate at the LATEST, in order to keep the amount of moisture entering the tubing
connections as low as possible.
NOTICE
– Reduced lifetime!
If Calorimetry Process Control Units and Air Pumps are powered ON while there is a leakage
in the sample line, this may reduce lifetime of devices.
•
Never remove collection vessels from the 4x Air Drying Unit. Drain off condensed water via
collection vessel tubing only.